Buy me a coffee
Home › FDA recalls › LVP software of the Ivenix Infusion System
CLASS I · SeriousFDA

Fresenius Kabi USA, LLC recalls LVP software of the Ivenix Infusion System

📅 May 6, 2026 · 📍 Nationwide
Recalling company
Fresenius Kabi USA, LLC
Product
LVP software of the Ivenix Infusion System (IIS); Product Code LVP-SW-0005.
Reason
Potential software anomaly that can cause the pump to report a false battery-health value (typically 69%) and trigger a battery-depletion alarm that will shut down the pump unless it is reconnected to AC mains power.
Classification
Class I – a reasonable probability that use will cause serious adverse health consequences or death
Recall date
May 6, 2026
Reported
Jun 10, 2026
Distribution
US Nationwide distribution in the states of CA, CO, FL, GA, GA, ID, IL, MD, MI, MN, MO, MS, NE, NJ, NV, OK, OR, SC, TX, VA, WA, WI.
Status
Ongoing
Quantity
32 systems
Lot / codes
Product Code LVP-SW-0005. UDI-DI: 00811505030122. Software Version 5.10.2
Company location
North Andover, MA
Recall type
Voluntary: Firm initiated
Recall number
Z-2223-2026
Agency
U.S. Food and Drug Administration (FDA)

View official FDA notice ↗ Search recall number Z-2223-2026 in the FDA Enforcement Report. Also see FDA recall press releases.

If you use this product, check the lot, NDC or model number against the official notice and follow its instructions. Do not stop a prescribed medicine or remove a medical device without talking to your doctor or pharmacist. Report problems to FDA MedWatch. FoodRecallNow summarizes public FDA data and is not medical advice.