Buy me a coffee
Home › FDA recalls › ARTIS Icono biplane
CLASS II · ModerateFDA

Siemens Medical Solutions USA, Inc recalls ARTIS Icono biplane

📅 Mar 9, 2026 · 📍 See notice
Recalling company
Siemens Medical Solutions USA, Inc
Product
ARTIS Icono biplane, floor and ceiling.: ARTIS icono biplane Model 11327600. ARTIS icono ceiling Model 11328100. ARTIS icono floor Model 11327700. interventional fluoroscopic x-ray system
Reason
During patient examinations, the system may sporadically display a 0 x-ray dose administered to the patient. This behavior has only been observed during fluoroscopy and/or single-frame acquisitions. Despite the reported x-ray dose showing 0, an actual x-ray dose is applied to the patient.
Classification
Class II – may cause temporary or medically reversible health consequences
Recall date
Mar 9, 2026
Reported
Jun 10, 2026
Distribution
US and Worldwide
Status
Ongoing
Lot / codes
ARTIS icono biplane UDI 4056869063317 ARTIS icono ceiling UDI 4056869295923 ARTIS icono floor UDI 4056869149325
Company location
Malvern, PA
Recall type
FDA Mandated
Recall number
Z-2206-2026
Agency
U.S. Food and Drug Administration (FDA)

View official FDA notice ↗ Search recall number Z-2206-2026 in the FDA Enforcement Report. Also see FDA recall press releases.

If you use this product, check the lot, NDC or model number against the official notice and follow its instructions. Do not stop a prescribed medicine or remove a medical device without talking to your doctor or pharmacist. Report problems to FDA MedWatch. FoodRecallNow summarizes public FDA data and is not medical advice.