Becton, Dickinson and Company recalls Swan-Ganz Catheter

📅 Apr 8, 2026 · 📍 Nationwide
Recalling company
Becton, Dickinson and Company
Product
Swan-Ganz Catheter, Models: 131F7/131F7P, 131F7J, 141F7, 151F7, 834F75, 096F6/096F6P, C144F7, 782F75M, 132F5, C146F7, 831F75/831F75P, TS105F5, 774F75, 777F8, AIQSGF8;
Reason
Catheters may, due to a manufacturing process and material changes, leak/break from the (blue) proximal injectate lumen hub, which may lead to infection, medication loss, and/or blood loss.
Classification
Class II – may cause temporary or medically reversible health consequences
Recall date
Apr 8, 2026
Reported
May 27, 2026
Distribution
Worldwide - US Nationwide distribution including in the states of TX, SC, CT, GA, VA, MO, NM, UT, NJ, CA, FL, PA, IN, TN, WA, KY, MD, WV, MI, MA, NC, MN, NY, AZ, AL, LA, CO, OK, HI, MS, IL, OH, NE, DC, KS, NV, WI, SD, IA, AR, ID, OR, WY, MT, RI, VT, DE, AK, NH, ND, ME, PR and the countries of Japan, Brazil, Colombia, Australia, New Zealand, Poland, Czech Republic, South Africa, Mauritius, Germany, Israel, Iran, Slovakia, Switzerland, Slovenia, Bulgaria, Cyprus, Saudi Arabia, United Arab Emirates (UAE), Kuwait, Malta, Bahrain, Qatar, Hungary, Morocco, Romania, Lebanon, Tunisia, Oman, Turkey, Austria, Belgium, Luxembourg, France, Netherlands, Italy, Spain, Canary Islands, Portugal, United Kingdom, Northern Ireland, Ireland, Denmark, Sweden, Iceland, Norway, Estonia, Finland, Greece, Chile, Puerto Rico, Canada.
Status
Ongoing
Quantity
589,223
Lot / codes
Model/UDI-DI/Lot(Expiration): 131F7P/00690103000412/65902900(8/1/2026), 65902901(8/3/2026), 65910728(8/2/2026), 65910729(7/29/2026),65910730(7/27/2026), 65910731(7/25/2026), 65910732(7/26/2026), 65910735(7/30/2026), 65910736(8/24/2026), 65910737(7/31/2026), 65916927(8/3/2026), 65916928(8/6/2026), 65916929(8/7/2026), 65916930(8/8/2026), 65916931(8/9/2026), 65916932(8/22/2026), 65916933(8/23/2026), 65944791(8/20/2026),65944792(8/21/2026), 65955857(9/9/2026), 65955858(9/17/2026), 65955859(9/26/2026), 65955860(9/27/2026), 65973739(9/5/2026), 65973740(9/6/2026), 65973741(9/7/2026), 65973742(9/10/2026), 65988571(9/28/2026), 65997478(9/16/2026), 65997479(9/17/2026), 65997481(9/19/2026), 65997482(9/20/2026), 66010504(9/10/2026), 66010505(9/11/2026), 66010506(9/12/2026), 66010507(9/13/2026), 66010508(9/14/2026), 66023074(9/30/2026), 66033088(10/5/2026), 66048930(10/9/2026), 66057629(10/6/2026), 66057630(10/8/2026), 66057944(10/15/2026), 66057945(10/17/2026), 66061552(10/24/2026), 66074999(10/4/2026), 66075318(11/19/2026), 66104892(10/21/2026), 66104893(10/22/2026), 66104894(10/23/2026), 66104895(11/13/2026), 66108277(11/15/2026), 66118638(11/22/2026), 66120310(10/29/2026), 66120312(10/30/2026), 66120313(10/31/2026), 66127027(11/15/2026), 66127028(11/20/2026), 66132570(11/21/2026), 66132571(11/25/2026), 66132572(11/27/2026), 66140447(11/2/2026), 66140448(11/5/2026), 66140449(11/6/2026), 66140450(11/7/2026), 66142269(11/28/2026), 66142274(11/29/2026), 66151008(12/20/2026), 66160820(11/3
Company location
Irvine, CA
Recall type
Voluntary: Firm initiated
Recall number
Z-2202-2026
Agency
U.S. Food and Drug Administration (FDA)

View official FDA notice ↗ Search recall number Z-2202-2026 in the FDA Enforcement Report. Also see FDA recall press releases.

If you use this product, check the lot, NDC or model number against the official notice and follow its instructions. Do not stop a prescribed medicine or remove a medical device without talking to your doctor or pharmacist. Report problems to FDA MedWatch. FoodRecallNow summarizes public FDA data and is not medical advice.