Buy me a coffee
Home › FDA recalls › Philips Azurion. System Code Description
CLASS II · ModerateFDA

PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. recalls Philips Azurion. System Code Description

📅 Mar 31, 2026 · 📍 Nationwide
Recalling company
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
Product
Philips Azurion. System Code Description (Model Number): Azurion 3 M12 (722221, 722229), Azurion 3 M15 (722222, 722230), Azurion 5 (722281), Azurion 5 M12 (722227, 722231), Azurion 5 M20 (722228, 722232, 722281), Azurion 7 (722282), Azurion 7 B12 (722225, 722235), Azurion 7 B20 (722068, 722226, 722236), Azurion 7 M12 (722078, 722223, 722233), Azurion 7 M20 (722079, 722224, 722234, 722282).
Reason
Potential for the bolts and plastic parts of the Cable Hose Carriers that hold the cable hose to the monitor ceiling suspension may loosen and/or break over time due to the forces applied when the monitor is moved or rotated.
Classification
Class II – may cause temporary or medically reversible health consequences
Recall date
Mar 31, 2026
Reported
May 20, 2026
Distribution
Worldwide distribution - US Nationwide and the country of Canada.
Status
Ongoing
Quantity
1,718 units
Lot / codes
1. System Code Description: Azurion 3 M12. Model Number [UDI]: 722221 [(01)00884838099203(21)181, (01)00884838099203(21)178, (01)00884838099203(21)93]; 722229 [(01)00884838116726(21)22, (01)00884838116726(21)9, (01)00884838116726(21)16, (01)00884838116726(21)15, (01)00884838116726(21)5, (01)00884838116726(21)3, (01)00884838116726(21)4, (01)00884838116726(21)6, (01)00884838116726(21)2, (01)00884838116726(21)8]. 2. System Code Description: Azurion 3 M15. Model Number [UDI]: 722222 [(01)00884838099210(21)358, (01)00884838099210(21)328, (01)00884838099210(21)389, (01)00884838099210(21)378, (01)00884838099210(21)243, (01)00884838099210(21)392, (01)00884838099210(21)396, (01)00884838099210(21)13, (01)00884838099210(21)379, (01)00884838099210(21)385, (01)00884838099210(21)376, (01)00884838099210(21)366, (01)00884838099210(21)365, (01)00884838099210(21)398, (01)00884838099210(21)274]; 722230 [(01)00884838116733(21)40, (01)00884838116733(21)2, (01)00884838116733(21)41, (01)00884838116733(21)15, (01)00884838116733(21)28, (01)00884838116733(21)38, (01)00884838116733(21)9, (01)00884838116733(21)27, (01)00884838116733(21)13]. 3. System Code Description: Azurion 5. Model Number [UDI]: 722281 [(01)00884838110564(21)401, (01)00884838110564(21)427, (01)00884838110564(21)382, (01)00884838110564(21)425]. 4. System Code Description: Azurion 5 M12. Model Number [UDI]: 722227 [(01)00884838099227(21)265, (01)00884838099227(21)360, (01)00884838099227(21)320, (01)00884838099227(21)338, (01)0088
Company location
Best
Recall type
Voluntary: Firm initiated
Recall number
Z-2170-2026
Agency
U.S. Food and Drug Administration (FDA)

View official FDA notice ↗ Search recall number Z-2170-2026 in the FDA Enforcement Report. Also see FDA recall press releases.

If you use this product, check the lot, NDC or model number against the official notice and follow its instructions. Do not stop a prescribed medicine or remove a medical device without talking to your doctor or pharmacist. Report problems to FDA MedWatch. FoodRecallNow summarizes public FDA data and is not medical advice.