PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. recalls Philips Azurion. System Code Description
- Recalling company
- PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
- Product
- Philips Azurion. System Code Description (Model Number): Azurion 3 M12 (722221, 722229), Azurion 3 M15 (722222, 722230), Azurion 5 (722281), Azurion 5 M12 (722227, 722231), Azurion 5 M20 (722228, 722232, 722281), Azurion 7 (722282), Azurion 7 B12 (722225, 722235), Azurion 7 B20 (722068, 722226, 722236), Azurion 7 M12 (722078, 722223, 722233), Azurion 7 M20 (722079, 722224, 722234, 722282).
- Reason
- Potential for the bolts and plastic parts of the Cable Hose Carriers that hold the cable hose to the monitor ceiling suspension may loosen and/or break over time due to the forces applied when the monitor is moved or rotated.
- Classification
- Class II – may cause temporary or medically reversible health consequences
- Recall date
- Mar 31, 2026
- Reported
- May 20, 2026
- Distribution
- Worldwide distribution - US Nationwide and the country of Canada.
- Status
- Ongoing
- Quantity
- 1,718 units
- Lot / codes
- 1. System Code Description: Azurion 3 M12. Model Number [UDI]: 722221 [(01)00884838099203(21)181, (01)00884838099203(21)178, (01)00884838099203(21)93]; 722229 [(01)00884838116726(21)22, (01)00884838116726(21)9, (01)00884838116726(21)16, (01)00884838116726(21)15, (01)00884838116726(21)5, (01)00884838116726(21)3, (01)00884838116726(21)4, (01)00884838116726(21)6, (01)00884838116726(21)2, (01)00884838116726(21)8]. 2. System Code Description: Azurion 3 M15. Model Number [UDI]: 722222 [(01)00884838099210(21)358, (01)00884838099210(21)328, (01)00884838099210(21)389, (01)00884838099210(21)378, (01)00884838099210(21)243, (01)00884838099210(21)392, (01)00884838099210(21)396, (01)00884838099210(21)13, (01)00884838099210(21)379, (01)00884838099210(21)385, (01)00884838099210(21)376, (01)00884838099210(21)366, (01)00884838099210(21)365, (01)00884838099210(21)398, (01)00884838099210(21)274]; 722230 [(01)00884838116733(21)40, (01)00884838116733(21)2, (01)00884838116733(21)41, (01)00884838116733(21)15, (01)00884838116733(21)28, (01)00884838116733(21)38, (01)00884838116733(21)9, (01)00884838116733(21)27, (01)00884838116733(21)13]. 3. System Code Description: Azurion 5. Model Number [UDI]: 722281 [(01)00884838110564(21)401, (01)00884838110564(21)427, (01)00884838110564(21)382, (01)00884838110564(21)425]. 4. System Code Description: Azurion 5 M12. Model Number [UDI]: 722227 [(01)00884838099227(21)265, (01)00884838099227(21)360, (01)00884838099227(21)320, (01)00884838099227(21)338, (01)0088
- Company location
- Best
- Recall type
- Voluntary: Firm initiated
- Recall number
- Z-2170-2026
- Agency
- U.S. Food and Drug Administration (FDA)
View official FDA notice ↗ Search recall number Z-2170-2026 in the FDA Enforcement Report. Also see FDA recall press releases.
If you use this product, check the lot, NDC or model number against the official notice and follow its instructions. Do not stop a prescribed medicine or remove a medical device without talking to your doctor or pharmacist. Report problems to FDA MedWatch. FoodRecallNow summarizes public FDA data and is not medical advice.