Home › FDA recalls › Medline Convenience Kits: 1) KIT OPHTHALMIC RUPTURE GLOBE
CLASS II · ModerateFDA

Medline Industries, LP recalls Medline Convenience Kits: 1) KIT OPHTHALMIC RUPTURE GLOBE

📅 Jan 7, 2026 · 📍 Nationwide
Recalling company
Medline Industries, LP
Product
Medline Convenience Kits: 1) KIT OPHTHALMIC RUPTURE GLOBE, Model Number: DYKMBNDL104; 2) KIT OPHTHALMIC RUPTURE GLOBE, Model Number: DYKMBNDL104A; 3) KIT OPHTHALMOLOGY CORNEAL TRAN, Model Number: DYKMBNDL159; 4) KIT OPHTHALMOLOGY CATARACT, Model Number: DYKMBNDL180; 5) KIT OPHTHALMOLOGY EYE MUSCLE, Model Number: DYKMBNDL42; 6) KIT SURG ONC PORT INSERTION, Model Number: DYKMBNDL43B; 7) KIT SURG ONC EXCISION BREAST, Model Number: DYKMBNDL84B; 8) KIT OPTHAMOLOGY BASIC, Model Number: DYKMBNDL98; 9) KIT OPTHAMOLOGY BASIC, Model Number: DYKMBNDL98A; 10) KIT OPTHAMOLOGY BASIC, Model Number: DYKMBNDL98B; 11) CAROMONT EYE TRAY-LF, Model Number: DYNJ17004L; 12) MINOR EYE PACK, Model Number: DYNJ17219C; 13) PEDS EYE/MUSCLE PACK, Model Number: DYNJ38791M; 14) EYE PACK, Model Number: DYNJ41207C; 15) OPHTHALMIC PACK, Model Number: DYNJ42043I; 16) BREAST RECONSTRUCTION-HOULE, Model Number: DYNJ44852J; 17) CUSTOM EYE PACK HA-LF, Model Number: DYNJ45585C; 18) EYE PLASTIC PACK, Model Number: DYNJ47859B; 19) EYE PLASTIC PACK, Model Number: DYNJ47859C; 20) CATARACT PACK, Model Number: DYNJ53019A; 21) CATARACT PACK, Model Number: DYNJ54869A; 22) VITRECTOMY PACK, Model Number: DYNJ55280B; 23) NDNW-EYE PACK, Model Number: DYNJ61779C; 24) EYE PK, Model Number: DYNJ62306; 25) VITRECTOMY, Model Number: DYNJ64220; 26) OSC BREAST PACK, Model Number: DYNJ65808C; 27) MILLS CATARACT PACK, Model Number: DYNJ67973A; 28) EYE PACK, Model Number: DYNJ68367B; 29) CATARACT PACK, Model Number: DYNJ68875D; 30) CATARACT PACK, Model Number: DYNJ68875F; 31) CATARACT PACK, Model Number: DYNJ68875G; 32) VITRECTOMY IP, Model Number: DYNJ84020; 33) PLASTIC PACK, Model Number: DYNJ86288; 34) OPHTHO PLASTIC, Model Number: DYNJ902256C; 35) RETINAL, Model Number: DYNJ907495; 36) CATARACT, Model Number: DYNJ909924; 37) CATARACT, Model Number: DYNJ909924A; 38) COMBO EYE, Model Number: DYNJ909925; 39) VITRECTOMY, Model Number: DYNJ909929; 40) RETINAL PK-LF, Model Number: DYNJC2364O
Reason
Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.
Classification
Class II – may cause temporary or medically reversible health consequences
Recall date
Jan 7, 2026
Reported
May 13, 2026
Distribution
Worldwide - US Nationwide distribution.
Status
Ongoing
Quantity
7760 units
Lot / codes
1) DYKMBNDL104, UDI-DI: 10193489478945(each), 40193489478946(case), Lot Number: 21ABV400; 2) DYKMBNDL104, UDI-DI: 10193489478945(each), 40193489478946(case), Lot Number: 21FBP706; 3) DYKMBNDL104, UDI-DI: 10193489478945(each), 40193489478946(case), Lot Number: 21IBV636; 4) DYKMBNDL104, UDI-DI: 10193489478945(each), 40193489478946(case), Lot Number: 22GBP895; 5) DYKMBNDL104A, UDI-DI: 10195327240622(each), 40195327240623(case), Lot Number: 22HBX475; 6) DYKMBNDL104A, UDI-DI: 10195327240622(each), 40195327240623(case), Lot Number: 22JBS899; 7) DYKMBNDL104A, UDI-DI: 10195327240622(each), 40195327240623(case), Lot Number: 22JBS907; 8) DYKMBNDL104A, UDI-DI: 10195327240622(each), 40195327240623(case), Lot Number: 22KBO773; 9) DYKMBNDL104A, UDI-DI: 10195327240622(each), 40195327240623(case), Lot Number: 23HBV380; 10) DYKMBNDL104A, UDI-DI: 10195327240622(each), 40195327240623(case), Lot Number: 24EBU650; 11) DYKMBNDL104A, UDI-DI: 10195327240622(each), 40195327240623(case), Lot Number: 25HBK503; 12) DYKMBNDL159, UDI-DI: 10193489478273(each), 40193489478274(case), Lot Number: 21ABQ655; 13) DYKMBNDL180, UDI-DI: 10193489478464(each), 40193489478465(case), Lot Number: 21ABT655; 14) DYKMBNDL180, UDI-DI: 10193489478464(each), 40193489478465(case), Lot Number: 21BBM375; 15) DYKMBNDL180, UDI-DI: 10193489478464(each), 40193489478465(case), Lot Number: 21FBP773; 16) DYKMBNDL42, UDI-DI: 10193489478884(each), 40193489478885(case), Lot Number: 21ABR620; 17) DYKMBNDL42, UDI-DI: 10193489478884(each), 4
Company location
Northfield, IL
Recall type
Voluntary: Firm initiated
Recall number
Z-2113-2026
Agency
U.S. Food and Drug Administration (FDA)

View official FDA notice ↗ Search recall number Z-2113-2026 in the FDA Enforcement Report. Also see FDA recall press releases.

If you use this product, check the lot, NDC or model number against the official notice and follow its instructions. Do not stop a prescribed medicine or remove a medical device without talking to your doctor or pharmacist. Report problems to FDA MedWatch. FoodRecallNow summarizes public FDA data and is not medical advice.