Buy me a coffee
Home › FDA recalls › See full list in RES
CLASS II · ModerateFDA

Medline Industries, LP recalls See full list in RES

📅 Jan 7, 2026 · 📍 Nationwide
Recalling company
Medline Industries, LP
Product
See full list in RES, exceeds character limit 1) GYN LAPAROSCOPY CDS, Model Number: CDS981504G; 2) BAP NORTH ROBOTIC SURGERY CDS, Model Number: CDS982407P; 3) ROBOTIC GYN CDS-1, Model Number: CDS982893I; 4) GYN LAPAROSCOPY CDS, Model Number: CDS983300J; 5) NHP ROBOTIC GYNE CDS, Model Number: CDS984243G; 6) ROBOTIC-LF, Model Number: CDS984543D; 7) ROBOTIC-LF, Model Number: CDS984543F; 8) ROBOTIC-LF, Model Number: CDS984543G; 9) ROBOTIC-LF, Model Number: CDS984543I; 10) GYN CDS, Model Number: CDS984860F; 11) GYN LAP PELVI, Model Number: CDS985423G; 12) GYN KIT, Model Number: DYKM1657A; 13) KIT GYN ONCOLOGY LAP, Model Number: DYKMBNDL100; 14) KIT ROBOTICS GYN, Model Number: DYKMBNDL109; 15) KIT ROBOTICS UROLOGY PROSTATE, Model Number: DYKMBNDL116D; 16) KIT ROBOTICS UROLOGY PROSTATE, Model Number: DYKMBNDL116F; 17) KIT GYN VAGINAL MAJOR, Model Number: DYKMBNDL119; 18) KIT GYN VAGINAL MAJOR, Model Number: DYKMBNDL119A; 19) KIT GYN ABDOMINAL MAJOR, Model Number: DYKMBNDL168; 20) KIT GYN ABDOMINAL MAJOR, Model Number: DYKMBNDL168A; 21) KIT THORACIC ROBOTIC LOBECTOMY, Model Number: DYKMBNDL184D; 22) KIT GYN ONCOLOGY LAP, Model Number: DYKMBNDL91;
Reason
Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.
Classification
Class II – may cause temporary or medically reversible health consequences
Recall date
Jan 7, 2026
Reported
May 13, 2026
Distribution
Worldwide - US Nationwide distribution.
Status
Ongoing
Quantity
15806 units
Lot / codes
1) CDS981504G, UDI-DI: 10195327157968(each), 40195327157969(case), Lot Number: 25BDB744; 2) CDS982407P, UDI-DI: 10195327110116(each), 40195327110117(case), Lot Number: 23EBE997; 3) CDS982407P, UDI-DI: 10195327110116(each), 40195327110117(case), Lot Number: 23GBL921; 4) CDS982407P, UDI-DI: 10195327110116(each), 40195327110117(case), Lot Number: 23GBO574; 5) CDS982407P, UDI-DI: 10195327110116(each), 40195327110117(case), Lot Number: 23IBC607; 6) CDS982407P, UDI-DI: 10195327110116(each), 40195327110117(case), Lot Number: 23JBQ095; 7) CDS982893I, UDI-DI: 10889942208050(each), 40889942208051(case), Lot Number: 24IBO240; 8) CDS983300J, UDI-DI: 10193489425048(each), 40193489425049(case), Lot Number: 21CDB781; 9) CDS984243G, UDI-DI: 10193489872521(each), 40193489872522(case), Lot Number: 21EBA117; 10) CDS984243G, UDI-DI: 10193489872521(each), 40193489872522(case), Lot Number: 21GBI171; 11) CDS984243G, UDI-DI: 10193489872521(each), 40193489872522(case), Lot Number: 21HBN807; 12) CDS984243G, UDI-DI: 10193489872521(each), 40193489872522(case), Lot Number: 21IBM804; 13) CDS984243G, UDI-DI: 10193489872521(each), 40193489872522(case), Lot Number: 21VBA272; 14) CDS984243G, UDI-DI: 10193489872521(each), 40193489872522(case), Lot Number: 22BBL450; 15) CDS984243G, UDI-DI: 10193489872521(each), 40193489872522(case), Lot Number: 22FBV462; 16) CDS984243G, UDI-DI: 10193489872521(each), 40193489872522(case), Lot Number: 22GBP114; 17) CDS984243G, UDI-DI: 10193489872521(each), 40193489872522(case), L
Company location
Northfield, IL
Recall type
Voluntary: Firm initiated
Recall number
Z-2109-2026
Agency
U.S. Food and Drug Administration (FDA)

View official FDA notice ↗ Search recall number Z-2109-2026 in the FDA Enforcement Report. Also see FDA recall press releases.

If you use this product, check the lot, NDC or model number against the official notice and follow its instructions. Do not stop a prescribed medicine or remove a medical device without talking to your doctor or pharmacist. Report problems to FDA MedWatch. FoodRecallNow summarizes public FDA data and is not medical advice.