Erbe Medical, LLC recalls ERBEFLO CleverCap CO2: Hybrid CO2 Tubing/Cap Set for Olympus Scopes & CO2 Sources
- Recalling company
- Erbe Medical, LLC
- Product
- ERBEFLO CleverCap CO2: Hybrid CO2 Tubing/Cap Set for Olympus Scopes & CO2 Sources, REF: 20325-239; Hybrid CO2 Tubing/Cap Set for Olympus Scopes & UCR, REF: 20325-240; Hybrid CO2 Tubing/Cap Set for Olympus Scopes, REF: 20325-206; Hybrid CO2 Tubing/Cap Set for Pentax Scopes, REF: 20325-207; ERBEFLO CleverCap Hybrid CO2 Tubing/Cap Set for Fujifilm Scopes, REF: 20325-248; Hybrid CO2 Tubing/Cap Set for Fujinon Scopes, REF: 20325-208
- Reason
- Devices providing sterile water/CO2/air to endoscopes were manufactured with a distal irrigation segment connector configuration that may be more susceptible to unintended water flow from distal tip under unsupported third-party connector use conditions, which may lead to water aspiration, which may progress to aspiration pneumonia, acute respiratory distress syndrome, or respiratory failure.
- Classification
- Class II – may cause temporary or medically reversible health consequences
- Recall date
- Mar 31, 2026
- Reported
- May 13, 2026
- Distribution
- US Nationwide distribution including in the states of IL, NY, CA, WA, UT, KY, PA, TX, AL, OH, FL, IN, NJ, NH, MI, CT, AR, MN, WI, AZ, CO, WV, MT, HI, TN, GA, VA, OK, RI, MO, NM, ID, SC, WY, KS, NC, MS, OR, MD, MA, DE, NE, AK, IA, DC, ME, LA, SD, ND, VT, PR, NV.
- Status
- Ongoing
- Quantity
- 260,962
- Lot / codes
- REF/UDI-DI/Lots: 20325-239/04065655000375/W4464506, W4464507, W4464512, W4464514, W4464515, W4464516, W4464517, W4465114, W4465115, W4465116, W4465702, W4465709, W4465991, W4466004, W4466266, W4466267, W4466268, W4466269, W4466271, W4466272, W4466273, W4466275, W4466278, W4466279, W4466813, W4466814, W4466815, W4466816, W4467057, W4467058, W4467059, W4467680, W4467681, W4468147, W4468148, W4468273, W4468274, W4468275, WO468776, WO468777, WO468778, WO468779, WO468780, WO469443, WO469444, WO469448, WO469449, WO469450, WO469451, WO469452, WO469933, WO469934, WO469947, WO469948, WO469949, WO469950, WO469951, WO469952, WO470768, WO470774, WO470775, WO470776, WO470777, WO470778, WO470896, WO470897, WO470913, WO470914, WO470915, WO470922, WO471236, WO471237, WO471238, WO471239, WO471240, WO471241, WO471242, WO473987, WO473988, WO473989, WO473991, WO475246, WO475247, WO475248, WO475249; 20325-240/04065655000405/W4466020, W4466021, WO468768, WO468769, WO468770, WO468771, WO469445, WO469446, WO469447, WO469936, WO469937, WO471244, WO471246, WO473170, WO473171, WO473172, WO473173, WO473174, WO473982, WO473983, WO473984, WO473985, WO474409, WO474410, WO474411, WO474412, WO474413, WO474414, WO475243; 20325-206/04065655000436/W4467039, W4467049, W4467678, W4467679, WO468582, WO470769, WO470770, WO470771, WO470772, WO470773, WO471076, WO471077, WO471078, WO471079, WO471390, WO478365, WO478366, WO478367; 20325-207/04065655000689/W4469939, W4469940, W4469941; 20325-248/04065655001006/W44
- Company location
- Tempe, AZ
- Recall type
- Voluntary: Firm initiated
- Recall number
- Z-2062-2026
- Agency
- U.S. Food and Drug Administration (FDA)
View official FDA notice ↗ Search recall number Z-2062-2026 in the FDA Enforcement Report. Also see FDA recall press releases.
If you use this product, check the lot, NDC or model number against the official notice and follow its instructions. Do not stop a prescribed medicine or remove a medical device without talking to your doctor or pharmacist. Report problems to FDA MedWatch. FoodRecallNow summarizes public FDA data and is not medical advice.