Buy me a coffee
Home › FDA recalls › Brand Name: Da Vinci X and Da Vinci Xi Product Name: da Vin
CLASS II · ModerateFDA

Intuitive Surgical, Inc. recalls Brand Name: Da Vinci X and Da Vinci Xi Product Name: da Vinci X

📅 Apr 2, 2026 · 📍 Nationwide
Recalling company
Intuitive Surgical, Inc.
Product
Brand Name: Da Vinci X and Da Vinci Xi Product Name: da Vinci X, Xi Surgical System Model/Catalog Number: IS4000; IS4200 Software Version: N/A Product Description: he da Vinci Xi, X Surgical systems each consist of three main components: the Surgeon Console (SC), the Vision Cart (VC), and the Patient Cart (PC). The PC is the operative component of the system, and its primary function is to support the instrument arms and camera arm. Each system have four Universal Surgical Manipulators (USMs) also known as arms, which allow the surgeon to control up to three instruments and an endoscope simultaneously to perform surgical tasks (e.g. cutting, suturing, tissue approximation) in a variety of procedures. As the surgeon moves the hand controls within the console, the PC moves the robotic instruments within the patient in a corresponding manner. The da Vinci system and its software are designed such that instruments motion precisely tracks the motion inputs from the hand controls during surgery. Component: The affected part based on this recall, 380647-xx, USM, is a sub component of da Vinci Surgical Instrument
Reason
Due to manipulator containing screws with in the arm sub-assembly that my be susceptible to breaking.
Classification
Class II – may cause temporary or medically reversible health consequences
Recall date
Apr 2, 2026
Reported
May 13, 2026
Distribution
Worldwide - U.S. Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, WA, WI, WV, and WY. The countries of Argentina, Australia, Austria, Belgium, Brazil, Canada, China, Croatia, Cyprus, Czech Republic, Denmark, Ecuador, Finland, France, Germany, Greece, Hong Kong, Hungary, India, Ireland, Israel, Italy, Japan, Luxembourg, Malaysia, Malta, Mexico, Netherlands, Norway, Poland, Portugal, Qatar, Romania, Saudi Arabia, Singapore, Slovakia, Slovenia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Turkey, United Arab Emirates, and United Kingdom.
Status
Ongoing
Quantity
2781
Lot / codes
Lot Code: UDI: 00886874114216 FDA Medical Device Listing Number: D531324 See attachment Affected Product List UDI: 00886874114216 FDA Medical Device Listing Number: D295931 See attachment Affected Product List
Company location
Sunnyvale, CA
Recall type
Voluntary: Firm initiated
Recall number
Z-2046-2026
Agency
U.S. Food and Drug Administration (FDA)

View official FDA notice ↗ Search recall number Z-2046-2026 in the FDA Enforcement Report. Also see FDA recall press releases.

If you use this product, check the lot, NDC or model number against the official notice and follow its instructions. Do not stop a prescribed medicine or remove a medical device without talking to your doctor or pharmacist. Report problems to FDA MedWatch. FoodRecallNow summarizes public FDA data and is not medical advice.