Medical Action Industries, Inc. 306 recalls Medical Action Industries Inc. Pack Cath BHH
- Recalling company
- Medical Action Industries, Inc. 306
- Product
- Medical Action Industries Inc. Pack Cath BHH, REF: BHCA49K, Sterile EO, Rx Only, QTY: 5/Case
- Reason
- Convenience kit contains a syringe in which the rotating adapter may unwind during use, resulting in a loose connection and/or full disconnection between the syringe and manifold"
- Classification
- Class I – a reasonable probability that use will cause serious adverse health consequences or death
- Recall date
- Mar 17, 2026
- Reported
- May 13, 2026
- Distribution
- US distribution to state of: AZ OUS: None
- Status
- Ongoing
- Quantity
- 2030 kits
- Lot / codes
- UDI:20809160490294/ Lot numbers: 342509, 346066, 338434, 339423, 337365, 335489, 336189, 336789
- Company location
- Arden, NC
- Recall type
- Voluntary: Firm initiated
- Recall number
- Z-2007-2026
- Agency
- U.S. Food and Drug Administration (FDA)
View official FDA notice ↗ Search recall number Z-2007-2026 in the FDA Enforcement Report. Also see FDA recall press releases.
If you use this product, check the lot, NDC or model number against the official notice and follow its instructions. Do not stop a prescribed medicine or remove a medical device without talking to your doctor or pharmacist. Report problems to FDA MedWatch. FoodRecallNow summarizes public FDA data and is not medical advice.