Abbott Point Of Care Inc. recalls i-STAT EG6+ cartridge
- Recalling company
- Abbott Point Of Care Inc.
- Product
- i-STAT EG6+ cartridge; List Number: 03P77-25;
- Reason
- Approximately 7.6% of cartridges from specific i-STAT EG7+ (list number 03P76-25), i-STAT EG6+ (list number 03P77-25) and i-STAT G3+ (list number 03P78-26) cartridge lots in the U.S. marketplace may report higher than expected PCO2 and lower than expected pH results due to a manufacturing issue. A falsely high pCO2 together with a falsely low pH makes the blood gas appear as if the patient is in severe respiratory acidosis. If the clinician believes the values are real, it may trigger unnecessary or even harmful interventions such as unnecessary ventilator changes, unnecessary medications, potential harm from overcorrection, or delays in identifying true problems.
- Classification
- Class II – may cause temporary or medically reversible health consequences
- Recall date
- Apr 1, 2026
- Reported
- May 6, 2026
- Distribution
- Worldwide distribution - US Nationwide and the countries of Canada, AUSTRIA, BANGLADESH, BELGIUM, FINLAND, FRANCE, GERMANY, IRELAND, ITALY, LUXEMBOURG, MAURITIUS, NETHERLANDS, NORWAY, POLAND, PORTUGAL, SPAIN, SWEDEN, SWITZERLAND, UNITED KINGDOM.
- Status
- Ongoing
- Quantity
- 30775 units (25625 US, 5150 OUS)
- Lot / codes
- List Number: 03P77-25; UDI-DI: 00054749000050; Lot number: N25223A;
- Company location
- Princeton, NJ
- Recall type
- Voluntary: Firm initiated
- Recall number
- Z-2000-2026
- Agency
- U.S. Food and Drug Administration (FDA)
View official FDA notice ↗ Search recall number Z-2000-2026 in the FDA Enforcement Report. Also see FDA recall press releases.
If you use this product, check the lot, NDC or model number against the official notice and follow its instructions. Do not stop a prescribed medicine or remove a medical device without talking to your doctor or pharmacist. Report problems to FDA MedWatch. FoodRecallNow summarizes public FDA data and is not medical advice.