Heraeus Medical GmbH (Dental Division) recalls Heraeus

📅 Mar 25, 2026 · 📍 Nationwide
Recalling company
Heraeus Medical GmbH (Dental Division)
Product
Heraeus, PALAMIX duo. Material Number: 66057897.
Reason
Based on long-term stability testing, it was determined that in units older than 1.5 years (17 months) the mixing rod may, in rare cases, come loose from its lock during the mixing process. Notification is to reduce the shelf life from 3 years to 1.5 years (17 months).
Classification
Class III – not likely to cause adverse health consequences
Recall date
Mar 25, 2026
Reported
May 6, 2026
Distribution
Worldwide distribution - US Nationwide and the countries of Aland Islands, Algeria, Australia, Austria, Belgium, Brazil , Chile, China, Cyprus, Czech Republic, Denmark, Ecuador, Finland, France, Germany, Greece, Hungary, India, Ireland, Israel, Italy, Kosovo, Lithuania, Luxembourg, Malaysia, Malta, Mexico, Netherlands, Nigeria, Norway, Peru, Philippines, Poland, Portugal, Puerto Rico, Rep. of Korea, Saudi Arabia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Thailand, United Kingdom, Uruguay, Utd.Arab Emir.;
Status
Ongoing
Quantity
1,125 units
Lot / codes
Material Number: 66057897; UDI-DI: 04260102133617; Batch: all batches with numbers below 520215;
Company location
Wehrheim
Recall type
Voluntary: Firm initiated
Recall number
Z-1998-2026
Agency
U.S. Food and Drug Administration (FDA)

View official FDA notice ↗ Search recall number Z-1998-2026 in the FDA Enforcement Report. Also see FDA recall press releases.

If you use this product, check the lot, NDC or model number against the official notice and follow its instructions. Do not stop a prescribed medicine or remove a medical device without talking to your doctor or pharmacist. Report problems to FDA MedWatch. FoodRecallNow summarizes public FDA data and is not medical advice.