Buy me a coffee
Home › FDA recalls › iLet Bionic Pancreas
CLASS II · ModerateFDA

Beta Bionics, Inc. recalls iLet Bionic Pancreas

📅 Mar 12, 2026 · 📍 Nationwide
Recalling company
Beta Bionics, Inc.
Product
iLet Bionic Pancreas, REF: BB1001
Reason
insulin pump software versions 1.4.2, and 1.4.3 issue with Dexcom Continuous Glucose Monitoring (CGM) G7 sensor may result in delayed CGM reading then switch to non-CGM dosing, leading to only basal/requested meal doses provided, no correction doses/insulin reductions/suspensions for high/low glucose unless blood glucose entered for sensor life or replacement. Potential health risks to users include severe hyperglycemia with and without DKA and severe hypoglycemia.
Classification
Class II – may cause temporary or medically reversible health consequences
Recall date
Mar 12, 2026
Reported
May 6, 2026
Distribution
US Nationwide distribution in the states of TX, PA, AL, MI, OH, FL, TN, GA.
Status
Ongoing
Quantity
1,080
Lot / codes
Software versions 1.4.3, and 1.4.2
Company location
Irvine, CA
Recall type
Voluntary: Firm initiated
Recall number
Z-1996-2026
Agency
U.S. Food and Drug Administration (FDA)

View official FDA notice ↗ Search recall number Z-1996-2026 in the FDA Enforcement Report. Also see FDA recall press releases.

If you use this product, check the lot, NDC or model number against the official notice and follow its instructions. Do not stop a prescribed medicine or remove a medical device without talking to your doctor or pharmacist. Report problems to FDA MedWatch. FoodRecallNow summarizes public FDA data and is not medical advice.