Buy me a coffee
Home › FDA recalls › BD GasPak EZ Campy Pouch System.
CLASS II · ModerateFDA

Becton Dickinson & Co. recalls BD GasPak EZ Campy Pouch System.

📅 Apr 7, 2026 · 📍 Nationwide
Recalling company
Becton Dickinson & Co.
Product
BD GasPak EZ Campy Pouch System. Catalog Number: 260685. for microbiologic use
Reason
BD identified an issue affecting certain lots of BD GasPak EZ CO2 and BD GasPak EZ Campy Gas Generating Pouch Systems, where gas generating sachets may produce carbon dioxide levels below specification.
Classification
Class II – may cause temporary or medically reversible health consequences
Recall date
Apr 7, 2026
Reported
May 6, 2026
Distribution
Worldwide distribution - US Nationwide and the country of Hong Kong.
Status
Ongoing
Quantity
2,200 units
Lot / codes
Catalog Number: 260685; UDI-DI: 30382902606858; Lot Number: 5329863.
Company location
Sparks, MD
Recall type
Voluntary: Firm initiated
Recall number
Z-1983-2026
Agency
U.S. Food and Drug Administration (FDA)

View official FDA notice ↗ Search recall number Z-1983-2026 in the FDA Enforcement Report. Also see FDA recall press releases.

If you use this product, check the lot, NDC or model number against the official notice and follow its instructions. Do not stop a prescribed medicine or remove a medical device without talking to your doctor or pharmacist. Report problems to FDA MedWatch. FoodRecallNow summarizes public FDA data and is not medical advice.