Buy me a coffee
Home › FDA recalls › Dimension Creatinine Flex reagent cartridge. Material Number
CLASS II · ModerateFDA

Siemens Healthcare Diagnostics, Inc. recalls Dimension Creatinine Flex reagent cartridge. Material Number: 10872079.

📅 Mar 18, 2026 · 📍 Nationwide
Recalling company
Siemens Healthcare Diagnostics, Inc.
Product
Dimension Creatinine Flex reagent cartridge. Material Number: 10872079.
Reason
Siemens Healthcare Diagnostics is recalling certain lots of Dimension Creatinine Flex reagent cartridge (Dimension CRE2) because a potential imprecision in the Dimension CRE2 quality control (QC) and patient sample results using lot numbers GA6307 and BA7005 on the Dimension System. This imprecision may lead to erroneously decreased or increased creatinine patient results
Classification
Class II – may cause temporary or medically reversible health consequences
Recall date
Mar 18, 2026
Reported
May 6, 2026
Distribution
Worldwide distribution - US Nationwide and the countries of Argentina, Australia, Austria, Bahamas, Brazil, Cambodia, Canada, Chile, China, Columbia, Costa Rica, Ecuador, Egypt, Germany, Honduras, India, Indonesia, Korea, Kuwait, Lebanon, Malaysia, Mexico, Nicaragua, Pakistan, Paraguay, Philippines, Qatar, Saudi Arabia, Serbia, Slovenia, South Africa, Sri Lanka, Taiwan, Thailand, U.A.E., Ukraine, Uruguay, Vietnam, Yemen.
Status
Ongoing
Quantity
18,080 units
Lot / codes
Material Number: 10872079. Model Number: DF33B. UDI: (01)00630414595009(10)GA6307(17)20261103, (01)00630414595009(10)BA7005(17)20270105. Lot Numbers: GA6307, BA7005.
Company location
Newark, DE
Recall type
Voluntary: Firm initiated
Recall number
Z-1980-2026
Agency
U.S. Food and Drug Administration (FDA)

View official FDA notice ↗ Search recall number Z-1980-2026 in the FDA Enforcement Report. Also see FDA recall press releases.

If you use this product, check the lot, NDC or model number against the official notice and follow its instructions. Do not stop a prescribed medicine or remove a medical device without talking to your doctor or pharmacist. Report problems to FDA MedWatch. FoodRecallNow summarizes public FDA data and is not medical advice.