Hologic, Inc recalls Brand Name: Genius" Review Station Product Name: Genius" Review Station Display
- Recalling company
- Hologic, Inc
- Product
- Brand Name: Genius" Review Station Product Name: Genius" Review Station Display (Barco Monitor) Model/Catalog Number: CMP-01669 Software Version: N/A - not software version specific Component: Yes, the Genius" Review Station is a component of the Genius" Digital Diagnostics System with the Genius" Cervical AI Algorithm cleared under DEN210035.
- Reason
- It was identified that certain devices were being operated outside the validated and FDA-cleared display configuration due to user modification of the manufacturer-installed settings. Modifications included disabling Barco Application Appearance Manager (AAM), altering calibration and compliance parameters, and adjusting additional display-related settings. This field correction is being conducted to restore affected system to the validated configuration and reinforce proper use in accordance with the Operator's Manual.
- Classification
- Class II – may cause temporary or medically reversible health consequences
- Recall date
- Mar 23, 2026
- Reported
- May 6, 2026
- Distribution
- Worldwide distribution - US Nationwide and the countries of Asia Pacific, Europe, and the Middle East.
- Status
- Ongoing
- Quantity
- 1200 units
- Lot / codes
- Model Number: CMP-01669; UDI-DI: 15420045514201; Serial Number: All;
- Company location
- Marlborough, MA
- Recall type
- Voluntary: Firm initiated
- Recall number
- Z-1949-2026
- Agency
- U.S. Food and Drug Administration (FDA)
View official FDA notice ↗ Search recall number Z-1949-2026 in the FDA Enforcement Report. Also see FDA recall press releases.
If you use this product, check the lot, NDC or model number against the official notice and follow its instructions. Do not stop a prescribed medicine or remove a medical device without talking to your doctor or pharmacist. Report problems to FDA MedWatch. FoodRecallNow summarizes public FDA data and is not medical advice.