Beckman Coulter, Inc. recalls Access Total T4 Calibrator
- Recalling company
- Beckman Coulter, Inc.
- Product
- Access Total T4 Calibrator, Catalog No. 33805
- Reason
- Beckman Coulter has identified that Access Total T4 Calibrator lots (PN 33805) exhibit a negative bias on DxI 600/800 instruments, with slopes outside product specifications. A false low patient result could cause a patient to undergo unnecessary additional diagnostic testing. A patient may be treated for a disorder not present or receive delayed treatment for hyperthyroidism. For a pregnant patient being monitored for thyroid hormone supplementation, incorrect treatment or dose may be used. A delayed patient results could cause repeat blood draw with associated complications.
- Classification
- Class II – may cause temporary or medically reversible health consequences
- Recall date
- Mar 16, 2026
- Reported
- Apr 29, 2026
- Distribution
- Worldwide distribution - US Nationwide and the countries of Argentina, Austria, Bangladesh, Bolivia (Plurinational State of), Bosnia and Herzegovina, Brazil, Canada, China, C¿te d'Ivoire, Croatia, Czechia, Denmark, Egypt, El Salvador, Ethiopia, Georgia, Germany, Greece, Hong Kong, India, Indonesia, Ireland, Italy, Japan, Kazakhstan, Korea, Republic of, Kosovo, Lebanon, Libya, Malawi, Mexico, Myanmar, Nepal, Netherlands, New Zealand, Nigeria, North Macedonia, Pakistan, Panama, Peru, Philippines, Poland, Portugal, Puerto Rico, Qatar, Romania, Russian Federation, Saudi Arabia, Serbia, Singapore, Slovakia, Somalia, Spain, Switzerland, Taiwan, Province of China, Thailand, Turkey, United Arab Emirates, Uzbekistan, Viet Nam, Yemen.
- Status
- Ongoing
- Quantity
- 2068 US, 2830 OUS
- Lot / codes
- UDI-DI 15099590225872 Lots 538729, 538815, 538898, 539107, 539243
- Company location
- Chaska, MN
- Recall type
- Voluntary: Firm initiated
- Recall number
- Z-1922-2026
- Agency
- U.S. Food and Drug Administration (FDA)
View official FDA notice ↗ Search recall number Z-1922-2026 in the FDA Enforcement Report. Also see FDA recall press releases.
If you use this product, check the lot, NDC or model number against the official notice and follow its instructions. Do not stop a prescribed medicine or remove a medical device without talking to your doctor or pharmacist. Report problems to FDA MedWatch. FoodRecallNow summarizes public FDA data and is not medical advice.