Home › FDA recalls › Philips Spectral CT on Rails.
CLASS II · ModerateFDA

PHILIPS MEDICAL SYSTEMS recalls Philips Spectral CT on Rails.

📅 Mar 7, 2026 · 📍 Nationwide
Recalling company
PHILIPS MEDICAL SYSTEMS
Product
Philips Spectral CT on Rails. Model Number: 728334.
Reason
Philips has identified three software issues: 1. During a continuous CT (CCT) scan, there is the potential that the Gantry could remain at the current scan position after pressing Go. 2. Potential where the message of Previous Surview Exists Select Previous Surview? <Yes> or <No> should display but does not appear. 3. After performing Surview and planning the Brain Helical acquisition by setting Brain Area DoseRight Index to increase the dose, there is the potential that the Define Head Area option in the context menu is grayed out.
Classification
Class II – may cause temporary or medically reversible health consequences
Recall date
Mar 7, 2026
Reported
Apr 22, 2026
Distribution
Worldwide - US Nationwide distribution in the states of Minnesota and the countries of France and Netherlands.
Status
Ongoing
Quantity
3 systems
Lot / codes
Model Number: 728334. UDI: (01)00884838103627(21)1005, (01)00884838103627(21)1006, (01)00884838103627(21)1007. Software Version Number: 5.1.X. Serial Numbers: 1005, 1006, 1007.
Company location
Cambridge, MA
Recall type
Voluntary: Firm initiated
Recall number
Z-1825-2026
Agency
U.S. Food and Drug Administration (FDA)

View official FDA notice ↗ Search recall number Z-1825-2026 in the FDA Enforcement Report. Also see FDA recall press releases.

If you use this product, check the lot, NDC or model number against the official notice and follow its instructions. Do not stop a prescribed medicine or remove a medical device without talking to your doctor or pharmacist. Report problems to FDA MedWatch. FoodRecallNow summarizes public FDA data and is not medical advice.