Buy me a coffee
Home › FDA recalls › Brand Name: Cellex Photopheresis System Product Name: CELLE
CLASS II · ModerateFDA

THERAKOS DEVELOPMENT LIMITED recalls Brand Name: Cellex Photopheresis System Product Name: CELLEX Photopheresis Kit

📅 May 16, 2025 · 📍 Nationwide
Recalling company
THERAKOS DEVELOPMENT LIMITED
Product
Brand Name: Cellex Photopheresis System Product Name: CELLEX Photopheresis Kit Model/Catalog Number: CLXUSA Software Version: Not Applicable Product Description: THERAKOS Photopheresis or extracorporeal photopheresis (ECP) is a photoimmune therapy where leukocytes are separated from whole blood via apheresis, combined with a photoactive drug (8-methoxypsoralen) and then exposed to ultraviolet A (UVA) light. All blood components, including the treated leukocytes, are returned to the patient. THERAKOS¿ Photopheresis utilizes the THERAKOS CELLEX System to combine cell separation and photoactivation into a single, closed and sterile circuit. The THERAKOS CELLEX Photopheresis System collects the buffy coat (leukocyte-enriched blood) from the patient in a discontinuous flow process and intermittently returns the remaining plasma and erythrocytes to the patient. The buffy coat is passed through the photoactivation module where the drug is activated with a precise amount of UVA light determined by the characteristics of the individual patient s buffy coat. After photoactivation, the buffy coat is immediately returned to the patient s bloodstream. Component: Not Applicable
Reason
Increased difficulty in installing the centrifuge bowl onto the centrifuge bowl holder. Improper installation may cause the centrifuge bowl to dislodge, resulting in a broken bowl during the prime cycle or treatment phase, leading to a delay in patient treatment and blood loss.
Classification
Class II – may cause temporary or medically reversible health consequences
Recall date
May 16, 2025
Reported
Apr 22, 2026
Distribution
US Nationwide distribution.
Status
Ongoing
Quantity
11,499 units
Lot / codes
Model/Catalog Number: CLXUSA; UDI-DI: 20705030200003; Lot Numbers: N301, N339, N340, N341, N342, N343, N344, N345, N346, N347; P302, P303, P304, P305, P306, P307, P308, P309, P310, P311, P312, P313, P314, P319, P320.
Company location
Blanchardstown, N/A
Recall type
Voluntary: Firm initiated
Recall number
Z-1822-2026
Agency
U.S. Food and Drug Administration (FDA)

View official FDA notice ↗ Search recall number Z-1822-2026 in the FDA Enforcement Report. Also see FDA recall press releases.

If you use this product, check the lot, NDC or model number against the official notice and follow its instructions. Do not stop a prescribed medicine or remove a medical device without talking to your doctor or pharmacist. Report problems to FDA MedWatch. FoodRecallNow summarizes public FDA data and is not medical advice.