Buy me a coffee
Home › FDA recalls › ARTIS Pheno VE30A and VE40A
CLASS II · ModerateFDA

Siemens Medical Solutions USA, Inc recalls ARTIS Pheno VE30A and VE40A

📅 Mar 12, 2026 · 📍 See notice
Recalling company
Siemens Medical Solutions USA, Inc
Product
ARTIS Pheno VE30A and VE40A, Model 10849000
Reason
During 3D acquisitions, lighter and darker patient images may be captured, which may result in less accurate 3D reconstruction. The variation of brightness is a result of a constant unregulated medium dose and the angular change of patient diameter caused by rotational acquisition. This may result in unintentional low-dose radiation exposure to a patient.
Classification
Class II – may cause temporary or medically reversible health consequences
Recall date
Mar 12, 2026
Reported
Apr 22, 2026
Distribution
U.S. and O.U.S.
Status
Ongoing
Quantity
73
Lot / codes
UDI 04056869046877
Company location
Malvern, PA
Recall type
FDA Mandated
Recall number
Z-1814-2026
Agency
U.S. Food and Drug Administration (FDA)

View official FDA notice ↗ Search recall number Z-1814-2026 in the FDA Enforcement Report. Also see FDA recall press releases.

If you use this product, check the lot, NDC or model number against the official notice and follow its instructions. Do not stop a prescribed medicine or remove a medical device without talking to your doctor or pharmacist. Report problems to FDA MedWatch. FoodRecallNow summarizes public FDA data and is not medical advice.