Beta Bionics, Inc. recalls iLet Ace Pump Kit REF: BB1001 iLet Dosing Decision Software
- Recalling company
- Beta Bionics, Inc.
- Product
- iLet Ace Pump Kit REF: BB1001 iLet Dosing Decision Software
- Reason
- it was found that in versions 1.3.7, 1.4.2, and 1.4.3, the Lock Screen and Limited Access Passcode Screen on the iLet graphical user interface (GUI) include certain icons displayed in the status bar that are active, thereby allowing the user to bypass those screens when those icons on the status bar are pressed, allowing unauthorized access while the device is in Limited Access Mode. A Health Risk associated with Limited access mode includes severe hypoglycemia due to unauthorized access to the iLet if someone were to make unauthorized meal announcements or stopped insulin delivery.
- Classification
- Class II – may cause temporary or medically reversible health consequences
- Recall date
- Aug 13, 2025
- Reported
- Apr 22, 2026
- Distribution
- US Nationwide distribution including in the states of AL, AR, AZ, CA, CO, CT, DE, FL, GA, ID, IL, IN, LA, MA, MD, ME, MI, ME, MI, MN, MO, MT, NC, NE, NH, NJ, NV, NY, OH, OK, OR, PA, RI, SC, TN, TX, UT, VA, WA, WI, WY AND WASHINGTON DC.
- Status
- Ongoing
- Quantity
- 15 units
- Lot / codes
- ILet Ace Pumps containing software version BOM000036 Rev C (version 1.4.2) or older.
- Company location
- Irvine, CA
- Recall type
- Voluntary: Firm initiated
- Recall number
- Z-1809-2026
- Agency
- U.S. Food and Drug Administration (FDA)
View official FDA notice ↗ Search recall number Z-1809-2026 in the FDA Enforcement Report. Also see FDA recall press releases.
If you use this product, check the lot, NDC or model number against the official notice and follow its instructions. Do not stop a prescribed medicine or remove a medical device without talking to your doctor or pharmacist. Report problems to FDA MedWatch. FoodRecallNow summarizes public FDA data and is not medical advice.