Buy me a coffee
Home › FDA recalls › Multi-parameter Command Module
CLASS II · ModerateFDA

Spacelabs Healthcare, Inc. recalls Multi-parameter Command Module

📅 Feb 27, 2026 · 📍 Nationwide
Recalling company
Spacelabs Healthcare, Inc.
Product
Multi-parameter Command Module, Model 91496, optioned with Masimo or Nellcor
Reason
Multi-parameter command module has a circuit board issue and when in Auto mode, Cardiac Output (CO) measurement may activate early, resulting in Bad Curve error with no CO measurement or Irregular Curve error with CO measurement not automatically excluded, which may result in inaccurate readings, or the need of additional readings, which may lead to delayed care and/or fluid overload.
Classification
Class II – may cause temporary or medically reversible health consequences
Recall date
Feb 27, 2026
Reported
Apr 15, 2026
Distribution
Worldwide - US Nationwide distribution including in the states of NY, NJ, AL, OK, IA, MT, WY, MN, TX, WI, PA, WA, OR, MI, CA, SC, ID, FL, MS, KS, SD, LA, OH, MO, NH, GA, MD, TN, AZ, PR and the countries of TAIWAN, CANADA, KUWAIT, FRANCE, SAUDI ARABIA, POLAND, PAKISTAN, MEXICO, ARGENTINA, PANAMA.
Status
Ongoing
Quantity
1790
Lot / codes
Serial Numbers: UDI-DI: 10841522106415, 10841522106422. 1496-223736, 1496-223737, 1496-223738, 1496-224375, 1496-224376, 1496-224400, 1496-224401, 1496-224402, 1496-224403, 1496-224404, 1496-224405, 1496-224406, 1496-223850, 1496-223851, 1496-223852, 1496-224448, 1496-224956, 1496-224957, 1496-224958, 1496-224959, 1496-224960, 1496-224961, 1496-224981, 1496-224258, 1496-223595, 1496-223596, 1496-223597, 1496-223598, 1496-223619, 1496-223620, 1496-223621, 1496-223622, 1496-223623, 1496-223624, 1496-224717, 1496-224718, 1496-223854, 1496-224373, 1496-224374, 1496-223894, 1496-223895, 1496-223896, 1496-223897, 1496-223898, 1496-223899, 1496-223900, 1496-223901, 1496-223902, 1496-223903, 1496-223904, 1496-223905, 1496-223906, 1496-223907, 1496-223918, 1496-225043, 1496-225044, 1496-225045, 1496-225046, 1496-225047, 1496-225048, 1496-225049, 1496-225050, 1496-225051, 1496-225052, 1496-225053, 1496-225054, 1496-225055, 1496-225056, 1496-225057, 1496-225058, 1496-225059, 1496-225060, 1496-225061, 1496-225062, 1496-225063, 1496-225064, 1496-225065, 1496-225066, 1496-225067, 1496-225068, 1496-224949, 1496-223760, 1496-223761, 1496-225154, 1496-224639, 1496-224719, 1496-224720, 1496-224721, 1496-225088, 1496-225089, 1496-224672, 1496-223908, 1496-223909, 1496-223910, 1496-223911, 1496-223912, 1496-223913, 1496-223914, 1496-223915, 1496-223916, 1496-223917, 1496-224637, 1496-224638, 1496-224752, 1496-224744, 1496-224745, 1496-224746, 1496-224747, 1496-224407, 1496-224408, 1496-224409, 1
Company location
Snoqualmie, WA
Recall type
Voluntary: Firm initiated
Recall number
Z-1793-2026
Agency
U.S. Food and Drug Administration (FDA)

View official FDA notice ↗ Search recall number Z-1793-2026 in the FDA Enforcement Report. Also see FDA recall press releases.

If you use this product, check the lot, NDC or model number against the official notice and follow its instructions. Do not stop a prescribed medicine or remove a medical device without talking to your doctor or pharmacist. Report problems to FDA MedWatch. FoodRecallNow summarizes public FDA data and is not medical advice.