Buy me a coffee
Home › FDA recalls › MEDLINE ANTERIOR HIP PACK DYNJ64672B
CLASS II · ModerateFDA

Medline Industries, LP recalls MEDLINE ANTERIOR HIP PACK DYNJ64672B

📅 Feb 16, 2026 · 📍 Nationwide
Recalling company
Medline Industries, LP
Product
MEDLINE ANTERIOR HIP PACK DYNJ64672B
Reason
Kits contain recalled Olympus biopsy valves. Olympus is investigating an increase in complaints associated with rubber fragment detachment in the slit of single use biopsy valves. Potential consequences associated with detachment of fragments from valve include foreign body in patient's tracheobronchial tree, which may require intervention for removal. Other potential consequences include inflammatory response, hypoxia, and prolongation of procedure.
Classification
Class II – may cause temporary or medically reversible health consequences
Recall date
Feb 16, 2026
Reported
Apr 8, 2026
Distribution
US Nationwide distribution in the states of CA, FL, IL, IN, KY, MA, MD, MT, NY, OH, PA, TN, TX.
Status
Ongoing
Quantity
14,379 kits total
Lot / codes
UDI-DI 10195327377496 (EA) 40195327377497 (CS) LOT 23LMD300
Company location
Northfield, IL
Recall type
Voluntary: Firm initiated
Recall number
Z-1711-2026
Agency
U.S. Food and Drug Administration (FDA)

View official FDA notice ↗ Search recall number Z-1711-2026 in the FDA Enforcement Report. Also see FDA recall press releases.

If you use this product, check the lot, NDC or model number against the official notice and follow its instructions. Do not stop a prescribed medicine or remove a medical device without talking to your doctor or pharmacist. Report problems to FDA MedWatch. FoodRecallNow summarizes public FDA data and is not medical advice.