Olympus Corporation of the Americas recalls Brand Name: Olympus SOLTIVE Premium SuperPulsed Laser System Product Name: SOLTIVE Premiu
- Recalling company
- Olympus Corporation of the Americas
- Product
- Brand Name: Olympus SOLTIVE Premium SuperPulsed Laser System Product Name: SOLTIVE Premium Laser System TFL-PLS Model/Catalog Number: TFL-PLS containing TFL-CPLU Product Description: An electricity powered device assembly which emits a high-power laser beam to vaporize/ablate soft tissue with moderate hemostasis, little charring, and a thin zone of necrosis. Component: No
- Reason
- Following a complaint investigation, Olympus identified a potential defect in the 24V power supply module in some SOLTIVE laser units which may cause the system to become inoperable. Additionally, smoke or a burning smell may occur. By design, the issue causing the smoke or burning smell would be contained within the internal laser console enclosure and would be self-extinguishing.
- Classification
- Class II – may cause temporary or medically reversible health consequences
- Recall date
- Feb 27, 2026
- Reported
- Apr 8, 2026
- Distribution
- Worldwide distribution - US Nationwide and the countries of Canada, Chile, Europe, Singapore, Australia, and Hong Kong.
- Status
- Ongoing
- Quantity
- 408 units
- Lot / codes
- Laser system: TFL-PLS; UDI: 00821925044111; Component Part: TFL-CPLU; UDI: 00821925044586; Serial #: MDUF220005, MDUF220005, MDUF220437, MDUF220437, MDUF220446, MDUF220446, MDUF190072, MDUF200324, MDUF210030, MDUF210204, MDUF210379, MDUF220001, MDUF220002, MDUF220003, MDUF220007, MDUF220008, MDUF220009, MDUF220010, MDUF220011, MDUF220012, MDUF220013, MDUF220014, MDUF220015, MDUF220016, MDUF220017, MDUF220018, MDUF220019, MDUF220020, MDUF220021, MDUF220022, MDUF220023, MDUF220024, MDUF220025, MDUF220026, MDUF220027, MDUF220028, MDUF220029, MDUF220030, MDUF220031, MDUF220032, MDUF220033, MDUF220034, MDUF220035, MDUF220036, MDUF220037, MDUF220038, MDUF220039, MDUF220040, MDUF220042, MDUF220043, MDUF220044, MDUF220045, MDUF220046, MDUF220047, MDUF220048, MDUF220049, MDUF220050, MDUF220051, MDUF220052, MDUF220053, MDUF220054, MDUF220055, MDUF220056, MDUF220057, MDUF220058, MDUF220060, MDUF220061, MDUF220062, MDUF220063, MDUF220064, MDUF220065, MDUF220066, MDUF220067, MDUF220069, MDUF220070, MDUF220071, MDUF220072, MDUF220074, MDUF220075, MDUF220076, MDUF220077, MDUF220078, MDUF220079, MDUF220080, MDUF220081, MDUF220082, MDUF220083, MDUF220084, MDUF220085, MDUF220086, MDUF220087, MDUF220088, MDUF220089, MDUF220090, MDUF220091, MDUF220092, MDUF220093, MDUF220094, MDUF220095, MDUF220096, MDUF220097, MDUF220098, MDUF220100, MDUF220101, MDUF220102, MDUF220103, MDUF220106, MDUF220109, MDUF220110, MDUF220111, MDUF220113, MDUF220114, MDUF220115, MDUF220116, MDUF220117, MDUF2
- Company location
- Center Valley, PA
- Recall type
- Voluntary: Firm initiated
- Recall number
- Z-1694-2026
- Agency
- U.S. Food and Drug Administration (FDA)
View official FDA notice ↗ Search recall number Z-1694-2026 in the FDA Enforcement Report. Also see FDA recall press releases.
If you use this product, check the lot, NDC or model number against the official notice and follow its instructions. Do not stop a prescribed medicine or remove a medical device without talking to your doctor or pharmacist. Report problems to FDA MedWatch. FoodRecallNow summarizes public FDA data and is not medical advice.