Buy me a coffee
Home › FDA recalls › BD Kiestra" ReadA
CLASS II · ModerateFDA

BD KIESTRA LAB AUTOMATION recalls BD Kiestra" ReadA

📅 Feb 6, 2026 · 📍 KS, MD, MN, NJ, TX, VA
Recalling company
BD KIESTRA LAB AUTOMATION
Product
BD Kiestra" ReadA; Catalog No.: 446948.
Reason
In certain situations, following a system reboot, one or more modules may have become unreachable via the remote connectivity interface. When this occurred, there was potential for delays in plate retrieval from the incubator while connectivity was restored. This issue was intermittent and did not affect test results, performance of the assay, or sample integrity.
Classification
Class II – may cause temporary or medically reversible health consequences
Recall date
Feb 6, 2026
Reported
Apr 8, 2026
Distribution
Domestic: KS, MD, MN, NJ, TX, VA; International: Australia, Austria, Belgium, Canada, China, Denmark, France, Germany, Hong Kong, Italy, Kuwait, Mexico, Morocco, Netherlands, Norway, Saudi Arabia, Spain, Sweden, Switzerland, Taiwan, United Kingdom;
Status
Ongoing
Quantity
82 units (7 US, 65 OUS)
Lot / codes
Catalog No.: 446948; UDI-DI: 00382904469481; Serial Numbers: All;
Company location
Drachten
Recall type
Voluntary: Firm initiated
Recall number
Z-1658-2026
Agency
U.S. Food and Drug Administration (FDA)

View official FDA notice ↗ Search recall number Z-1658-2026 in the FDA Enforcement Report. Also see FDA recall press releases.

If you use this product, check the lot, NDC or model number against the official notice and follow its instructions. Do not stop a prescribed medicine or remove a medical device without talking to your doctor or pharmacist. Report problems to FDA MedWatch. FoodRecallNow summarizes public FDA data and is not medical advice.