Olympus Corporation of the Americas recalls Brand Name: Olympus PKS Cutting Forceps w/Cord
- Recalling company
- Olympus Corporation of the Americas
- Product
- Brand Name: Olympus PKS Cutting Forceps w/Cord, 5mm 24cm Product Name: PKS Cutting Forceps Model/Catalog Number: 920000PK Product Description: The PKS Cutting Forceps are intended to be passed through a 5mm cannula. Coagulation is achieved using electrosurgical energy under visualization. The device is intended to be used with the PKS electrical generator.
- Reason
- Olympus identified that the Cutting Forceps contain components for which the supplier did not adequately validate the welding process. Defective welds can result in the Cutting Forceps' jaw breaking during clinical use.
- Classification
- Class II – may cause temporary or medically reversible health consequences
- Recall date
- Feb 25, 2026
- Reported
- Apr 1, 2026
- Distribution
- Worldwide - US Nationwide and the countries of Canada, Mexico, EU, Australia, China, Singapore, Korea, Japan.
- Status
- Ongoing
- Quantity
- 7 units
- Lot / codes
- Model: 920000PK; UDI: 00821925038080; Lot Numbers: All unexpired;
- Company location
- Center Valley, PA
- Recall type
- Voluntary: Firm initiated
- Recall number
- Z-1643-2026
- Agency
- U.S. Food and Drug Administration (FDA)
View official FDA notice ↗ Search recall number Z-1643-2026 in the FDA Enforcement Report. Also see FDA recall press releases.
If you use this product, check the lot, NDC or model number against the official notice and follow its instructions. Do not stop a prescribed medicine or remove a medical device without talking to your doctor or pharmacist. Report problems to FDA MedWatch. FoodRecallNow summarizes public FDA data and is not medical advice.