Buy me a coffee
Home › FDA recalls › Blue Ventilator Adapter Module
CLASS I · SeriousFDA

Baxter Healthcare Corporation recalls Blue Ventilator Adapter Module

📅 Feb 18, 2026 · 📍 FL
Recalling company
Baxter Healthcare Corporation
Product
Blue Ventilator Adapter Module, single-patient use circuit, Product Code M07937. Accessory to Baxter Volara Respiratory Therapy system.
Reason
Urgent Medical Device Correction for the Volara system single-patient use circuit due to the potential that patients or caregivers may be unaware of a possible decrease in oxygen levels (oxygen desaturation), or potential lung tissue injury from overexpansion (barotrauma) when the Volara device is used in the home care environment. This correction is providing the home caregiver specific instructions for use.
Classification
Class I – a reasonable probability that use will cause serious adverse health consequences or death
Recall date
Feb 18, 2026
Reported
Apr 8, 2026
Distribution
Florida
Status
Ongoing
Quantity
19 units
Lot / codes
UDI/DI number 00887761984622
Company location
Deerfield, IL
Recall type
Voluntary: Firm initiated
Recall number
Z-1637-2026
Agency
U.S. Food and Drug Administration (FDA)

View official FDA notice ↗ Search recall number Z-1637-2026 in the FDA Enforcement Report. Also see FDA recall press releases.

If you use this product, check the lot, NDC or model number against the official notice and follow its instructions. Do not stop a prescribed medicine or remove a medical device without talking to your doctor or pharmacist. Report problems to FDA MedWatch. FoodRecallNow summarizes public FDA data and is not medical advice.