I.T.S. GmbH recalls I.T.S. LRS
- Recalling company
- I.T.S. GmbH
- Product
- I.T.S. LRS (Locking Reconstruction System) with the below descriptions and corresponding article numbers. 1. DFL (Distal Femur Locking) Distal Femur Plate; Article Numbers: 21301-5, 21301-7, 21301-9, 21301-13, 21302-5, 21302-7, 21302-9, 21302-13. 2. PTL (Proximal Lateral Tibia Locking) Proximal Lateral Tibia Plate; Article Numbers: 21321-7, 21321-12, 21322-7, 21322-12. 3.PTL (Proximal Lateral Tibia Locking) Proximal Medial Tibia Plate; Article Numbers: 21325-4, 21325-7.
- Reason
- Updated MRI safety testing demonstrates higher RF-induced temperature increases under certain MRI conditions than previously reflected in the Instructions for Use (IFU).
- Classification
- Class II – may cause temporary or medically reversible health consequences
- Recall date
- Jan 23, 2026
- Reported
- Apr 1, 2026
- Distribution
- US Nationwide distribution in the states of AL, AZ, BVI, CA, CO, FL, GA, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NE, NJ, NM, NV, OH, OK, PA, SC, TX, USVI, WV, W.
- Status
- Ongoing
- Quantity
- 176 units
- Lot / codes
- 1. Article Number (Lot Numbers): 21301-5 (26/F62569), 21301-7 (26/F62570), 21301-9 (26/F67518), 21301-13 (26/F62568, 26/F67815), 21302-5 (26/F67520), 21302-7 (26/F67821), 21302-9 (26/F62572, 26/F67822), 21302-13 (26/F62571, 26/F67519, 26/F67820). 2. Article Number (Lot Numbers): 21321-7 (26/F54955, 26/F62576), 21321-12 (26/F55883), 21322-7 (26/F55890, 26/F62578), 21322-12 (26/F62577). 3. Article Number (Lot Numbers): 21325-4 (26/F70528, 679/022433, 679/032217, 679/102240, 679/102242, 679/102303), 21325-7 (679/102241, 679/102304).
- Company location
- Lasnitzhohe
- Recall type
- Voluntary: Firm initiated
- Recall number
- Z-1603-2026
- Agency
- U.S. Food and Drug Administration (FDA)
View official FDA notice ↗ Search recall number Z-1603-2026 in the FDA Enforcement Report. Also see FDA recall press releases.
If you use this product, check the lot, NDC or model number against the official notice and follow its instructions. Do not stop a prescribed medicine or remove a medical device without talking to your doctor or pharmacist. Report problems to FDA MedWatch. FoodRecallNow summarizes public FDA data and is not medical advice.