Buy me a coffee
Home › FDA recalls › Pelvic Reconstruction System
CLASS II · ModerateFDA

I.T.S. GmbH recalls Pelvic Reconstruction System

📅 Jan 23, 2026 · 📍 Nationwide
Recalling company
I.T.S. GmbH
Product
Pelvic Reconstruction System (PRS) with the below descriptions and corresponding article numbers. 1. Fixation Screw; Article Number: 70312. 2. Spike short; Article Number: 70314.
Reason
Updated MRI safety testing demonstrates higher RF-induced temperature increases under certain MRI conditions than previously reflected in the Instructions for Use (IFU).
Classification
Class II – may cause temporary or medically reversible health consequences
Recall date
Jan 23, 2026
Reported
Apr 1, 2026
Distribution
US Nationwide distribution in the states of AL, AZ, BVI, CA, CO, FL, GA, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NE, NJ, NM, NV, OH, OK, PA, SC, TX, USVI, WV, W.
Status
Ongoing
Quantity
11 units
Lot / codes
1. Article Number (Lot Number): 70312 (26/F65133). 2. Article Number (Lot Number): 70314 (86/619736).
Company location
Lasnitzhohe
Recall type
Voluntary: Firm initiated
Recall number
Z-1600-2026
Agency
U.S. Food and Drug Administration (FDA)

View official FDA notice ↗ Search recall number Z-1600-2026 in the FDA Enforcement Report. Also see FDA recall press releases.

If you use this product, check the lot, NDC or model number against the official notice and follow its instructions. Do not stop a prescribed medicine or remove a medical device without talking to your doctor or pharmacist. Report problems to FDA MedWatch. FoodRecallNow summarizes public FDA data and is not medical advice.