Automated Impella Controller (AIC), labeled as the following with corresponding Product Codes: 1. Impella Controller, Packaged, CA; Product Code: 0042-0000-CA. 2. Impella Controller, Packaged, EU; Product Code: 0042-0000-EU. 3. Impella Controller, Packaged, UK; Product Code: 0042-0000-UK. 4. Impella Controller, Packaged, US; Product Code: 0042-0000-US. 5. Impella Optical Controller, Packaged, EU; Product Code: 0042-0010-EU. 6. Impella Optical Controller, Packaged, UK; Product Code: 0042-0010-UK. 7. Optical AIC w/Impella Connect, Pack'd, CA; Product Code: 0042-0040-CA. 8. Optical AIC w/Impella Connect, Pack'd, EU; Product Code: 0042-0040-EU. 9. Optical, AIC, Impella Connect, Pkgd, AU; Product Code: 0042-0040-AU.
Reason
Potential that the "Purge System Blocked" alarm display on the Automated Impella Controller (AIC) could be delayed when using first generation Impella 5.5 (0550-0008*) pumps.
Classification
Class I – a reasonable probability that use will cause serious adverse health consequences or death
Recall date
Feb 16, 2026
Reported
Apr 8, 2026
Distribution
International distribution in the country of Australia, Canada, France, Germany, Kuwait.
Status
Ongoing
Quantity
62 units
Lot / codes
All products below with software version 8.2.2 through 8.5 are affected. 1. Product Code: 0042-0000-CA; GTIN: 00813502011272; Serial Numbers: IC2035, IC9416. 2. Product Code: 0042-0000-EU; GTIN: 00813502011289; Serial Numbers: IC1677, IC3049, IC3099, IC3102, IC4358, IC4329, IC5774, IC5757, IC6580, IC6581. 3. Product Code: 0042-0000-UK; GTIN: 00813502011296; Serial Numbers: IC2023, IC1910, IC2704, IC3325, IC3312. 4. Product Code: 0042-0000-US; GTIN: 00813502010022; Serial Numbers: IC1140. 5. Product Code: 0042-0010-EU; GTIN: 00813502010961; Serial Numbers: IC5403, IC5404, IC5635, IC5636, IC5666, IC5667, IC5801, IC5904, IC5918, IC5921, IC6421, IC6422, IC8683. 6. Product Code: 0042-0010-UK; GTIN: 00813502010978; Serial Numbers: IC8101, IC8102, IC8103, IC8104, IC8544, IC9446, IC9449, IC11440. 7. Product Code: 0042-0040-CA; GTIN: 00813502012958; Serial Numbers: IC10306, IC10312, IC12604, IC12891, IC12895, IC12892, IC12605, IC12894. 8. Product Code: 0042-0040-EU; GTIN: 00813502012217; Serial Numbers: IC9014, IC9022, IC9429, IC10408, IC10425. 9. Product Code: 0042-0040-AU; GTIN: 00813502012972; Serial Numbers: IC10163, IC10387, IC11546, IC11566, IC11490, IC11544, IC12008, IC12054, IC11472, IC12151.
Company location
Danvers, MA
Recall type
Voluntary: Firm initiated
Recall number
Z-1591-2026
Agency
U.S. Food and Drug Administration (FDA)

View official FDA notice ↗ Search recall number Z-1591-2026 in the FDA Enforcement Report. Also see FDA recall press releases.

If you use this product, check the lot, NDC or model number against the official notice and follow its instructions. Do not stop a prescribed medicine or remove a medical device without talking to your doctor or pharmacist. Report problems to FDA MedWatch. FoodRecallNow summarizes public FDA data and is not medical advice.