Merit Medical Systems, Inc. recalls ProGuide Chronic Dialysis Catheters
- Recalling company
- Merit Medical Systems, Inc.
- Product
- ProGuide Chronic Dialysis Catheters, REF: DC01455550/C, DC21452419/C, DC21452419-NE5/C, DC21452823-NE5/C, DC21454035/C, DC21455550/C
- Reason
- 16F dual-valved splittable sheath introducer due to a design defect, may not split as intended, which may result in hemorrhage, foreign bodies, procedure delay, embolization/thrombosis, impaired catheter function, loss of vessel for future vascular access.
- Classification
- Class I – a reasonable probability that use will cause serious adverse health consequences or death
- Recall date
- Feb 13, 2026
- Reported
- Apr 1, 2026
- Distribution
- Worldwide distribution: US (nationwide) to states of: TX, LA, OR, NM, MI, FL, CO, CA, GA, NC, AK, NY, PA, WI, NJ, VA, AZ, IN, KY, MO, DE, SC, MT, NE, WV, OH, MA, MN, CT, MD, KS, WA, TN, ID, UT, DC, MS, IL, OK, AL, RI, AR, NV and OUS (International) to countries of : Canada, Spain, United Arab Emirates, Brazil, Jordan, Iraq, France, Indonesia, Malaysia, Thailand, Slovakia, Colombia, Saudi Arabia, Belize, Hong Kong, Mexico, Austria, Belgium, Chile, Germany, Sweden, Ireland, Poland, Latvia, Italy, South Africa, Switzerland, Bulgaria, Peru, Argentina, Qatar, Greece, Georgia, United Kingdom, Brunei Darussalam, Ecuador, Panama, Netherlands, Venezuela, Bolivarian Republic of, Costa Rica, Viet Nam, Cayman Islands, Iran, Islamic Republic of, Australia, Nicaragua, Dominican Republic, Bolivia, Plurinational State of, El Salvador, Guatemala, Honduras, Japan
- Status
- Ongoing
- Quantity
- 131
- Lot / codes
- REF:UDI-DI/Lot(Expiration): DC01455550/C:00884450697870/H2600169; DC21452419/C:00884450697900/H2659228: DC21452419-NE5/C:00884450697894/H3249278): DC21452823/C:00884450697924/H2600164, H2628890; DC21452823-NE5/C:00884450697917/H2990807: DC21453227/C:00884450697948/H2630602; DC21454035/C:00884450697962/H2628887: DC21455550/C:00884450697979/H2630553, H2659229
- Company location
- South Jordan, UT
- Recall type
- Voluntary: Firm initiated
- Recall number
- Z-1576-2026
- Agency
- U.S. Food and Drug Administration (FDA)
View official FDA notice ↗ Search recall number Z-1576-2026 in the FDA Enforcement Report. Also see FDA recall press releases.
If you use this product, check the lot, NDC or model number against the official notice and follow its instructions. Do not stop a prescribed medicine or remove a medical device without talking to your doctor or pharmacist. Report problems to FDA MedWatch. FoodRecallNow summarizes public FDA data and is not medical advice.