Buy me a coffee
Home › FDA recalls › Brand Name: Welch Allyn
CLASS II · ModerateFDA

Baxter Healthcare Corporation recalls Brand Name: Welch Allyn

📅 Feb 18, 2026 · 📍 Nationwide
Recalling company
Baxter Healthcare Corporation
Product
Brand Name: Welch Allyn, Inc. Product Name: Welch Allyn Spot Vision Screener VS100 Model/Catalog Number: VS100 Software Version: Software version 3.2.0.1 Product Description: Welch Allyn Spot Vision Screener VS100, FDA product code name: Refractometer, ophthalmic One unit per package
Reason
Display screen may unintentionally flicker due to a software issue. Exposure to flickering lights may induce photosensitive seizure activity in susceptible individuals.
Classification
Class II – may cause temporary or medically reversible health consequences
Recall date
Feb 18, 2026
Reported
Mar 25, 2026
Distribution
Worldwide - US Nationwide distribution in the states of AZ, CO, IL, IN, KS, MA, NE, NJ, NM, NY, PA, SC, TX, VA and the countries of Colombia and India.
Status
Ongoing
Quantity
59
Lot / codes
Lot Code: " Product Code: HKO " Version or Model: 901029 " Catalog Number: VS100 " Company Name: WELCH ALLYN, INC. " Primary DI Number: 00732094214154 " Packaging DI Number: 00732094212228 " Device Description: VS100 UDI BASE UNIT " Software version: 3.2.0.1 " Serial Numbers: Refer to attached Affected Product Table
Company location
Deerfield, IL
Recall type
Voluntary: Firm initiated
Recall number
Z-1573-2026
Agency
U.S. Food and Drug Administration (FDA)

View official FDA notice ↗ Search recall number Z-1573-2026 in the FDA Enforcement Report. Also see FDA recall press releases.

If you use this product, check the lot, NDC or model number against the official notice and follow its instructions. Do not stop a prescribed medicine or remove a medical device without talking to your doctor or pharmacist. Report problems to FDA MedWatch. FoodRecallNow summarizes public FDA data and is not medical advice.