Howmedica Osteonics Corp. recalls MICS3 Angled Sagittal Saw Attachment

📅 Feb 2, 2026 · 📍 Nationwide
Recalling company
Howmedica Osteonics Corp.
Product
MICS3 Angled Sagittal Saw Attachment; Part Number: 210490
Reason
A potential issue was identified with the torque strength on the external screws of the MICS3 Angled Sagittal Saw Attachment. These screws, which secure the cover to the Angled Saw Housing, may loosen during use and detach from the attachment.
Classification
Class II – may cause temporary or medically reversible health consequences
Recall date
Feb 2, 2026
Reported
Mar 25, 2026
Distribution
Domestic: Nationwide Distribution; International: Australia, Germany, Belgium, France, Spain UK, New Zealand;
Status
Ongoing
Quantity
1098 units
Lot / codes
Part Number: 210490; UDI: (01)07613327523874(10)TU6563205RP; Lot numbers: ALL;
Company location
Mahwah, NJ
Recall type
Voluntary: Firm initiated
Recall number
Z-1572-2026
Agency
U.S. Food and Drug Administration (FDA)

View official FDA notice ↗ Search recall number Z-1572-2026 in the FDA Enforcement Report. Also see FDA recall press releases.

If you use this product, check the lot, NDC or model number against the official notice and follow its instructions. Do not stop a prescribed medicine or remove a medical device without talking to your doctor or pharmacist. Report problems to FDA MedWatch. FoodRecallNow summarizes public FDA data and is not medical advice.