Buy me a coffee
Home › FDA recalls › HealthCast "Vital Sync" Remote Patient Monitoring System whi
CLASS II · ModerateFDA

Covidien LLC recalls HealthCast "Vital Sync" Remote Patient Monitoring System which consists of: Product Numbe

📅 Jan 30, 2026 · 📍 Nationwide
Recalling company
Covidien LLC
Product
HealthCast "Vital Sync" Remote Patient Monitoring System which consists of: Product Number: VSLBASE03P / VITALSYNCSW03 Model / UDI-DI codes: Vital Sync 3.3.1 / 10884521844933 Vital Sync 3.4.0 / 10884521844940 Vital Sync 3.4.1 / 10884521847569 (OUS Only) Vital Sync 3.4.2 / 10884521856127 Product Number: VSLBASE04P / VITALSYNCSW04 Model / UDI-DI codes: Vital Sync 4.0.0 / 10884521849358 Vital Sync 4.0.2 / 10884521856134 The Vital Sync" Informatics Manager is software that is intended to route and store medical device data and device diagnostic information from supported devices to the Virtual Patient Monitoring Platform, 3rd Party Annunciation Systems, Electronic Medical Record (EMR) and Clinical Information System (CIS).
Reason
Due to complaints and investigations stating that alarms from the primary patient bedside monitor was not alarming/being transmitted and received on the remote patient monitoring system.
Classification
Class II – may cause temporary or medically reversible health consequences
Recall date
Jan 30, 2026
Reported
Mar 18, 2026
Distribution
Worldwide - U.S. Nationwide distribution in the states of AK, AL, AZ, CA, FL, GA, IL, IN, KY, LA, MI, MN, MS, NC, NY, OH, OK, OR, SC, TN, TX, and VA. The country of Spain.
Status
Ongoing
Quantity
78 systems
Lot / codes
HealthCast "Vital Sync" Remote Patient Monitoring System which consists of: CFN: VSLBASE03P / VITALSYNCSW03 UDI-DI (GTIN); 10884521844933 Serial Number: Vital Sync 3.3.1 CFN: VSLBASE03P / VITALSYNCSW03 UDI-DI (GTIN): 10884521844940 Serial Number: Vital Sync 3.4.0 CFN: VSLBASE03P / VITALSYNCSW03 (OUS only) UDI-DI (GTIN): 10884521847569 Serial Number: Vital Sync 3.4.1 CFN: VSLBASE03P / VITALSYNCSW03 UDI-DI (GTIN): 10884521856127 Serial Number: Vital Sync 3.4.2 CFN: VSLBASE04P / VITALSYNCSW04 UDI-DI (GTIN): 10884521849358 Serial Number: Vital Sync 4.0.1 CFN: VSLBASE04P / VITALSYNCSW04 UDI-DI (GTIN): 10884521856134 Serial Number: Vital Sync 4.0.2
Company location
Boulder, CO
Recall type
Voluntary: Firm initiated
Recall number
Z-1557-2026
Agency
U.S. Food and Drug Administration (FDA)

View official FDA notice ↗ Search recall number Z-1557-2026 in the FDA Enforcement Report. Also see FDA recall press releases.

If you use this product, check the lot, NDC or model number against the official notice and follow its instructions. Do not stop a prescribed medicine or remove a medical device without talking to your doctor or pharmacist. Report problems to FDA MedWatch. FoodRecallNow summarizes public FDA data and is not medical advice.