Boston Scientific Corporation recalls EndoVive Safety PEG Kit
- Recalling company
- Boston Scientific Corporation
- Product
- EndoVive Safety PEG Kit; Outer Box Material Number (UPN): M00509001; Inner Box Material Number (UPN): M00509000;
- Reason
- Recall of kits which utilized ChloraPrep Triple Swabsticks from BD. Open or compromised seals on the sterile swabstick packaging, which could result in loss of sterility or drying of the antiseptic solution. This component is placed outside of the sterile barrier of the tray manufactured by Boston Scientific, which contains the majority of the kit components. As such, inclusion of the affected ChloraPrep Triple Swabsticks does not degrade the integrity of the other components within the PEG kit.
- Classification
- Class II – may cause temporary or medically reversible health consequences
- Recall date
- Feb 10, 2026
- Reported
- Mar 18, 2026
- Distribution
- US Nationwide distribution.
- Status
- Ongoing
- Quantity
- 806 units
- Lot / codes
- Outer Box Material Number (UPN): M00509001; Inner Box Material Number (UPN): M00509000; Outer Box UDI: 08714729880394; Inner Box UDI: 08714729880387; Outer Box Lot#s: 37435463, 37522541, 37528545, 37546847, 37546850, 37546849, 37550500, 37592036, 37732292; Inner Box Lot#s: 37435462, 37514809, 37525231, 37536563, 37536564, 37536562, 37546840, 37583642, 37731897;
- Company location
- Marlborough, MA
- Recall type
- Voluntary: Firm initiated
- Recall number
- Z-1551-2026
- Agency
- U.S. Food and Drug Administration (FDA)
View official FDA notice ↗ Search recall number Z-1551-2026 in the FDA Enforcement Report. Also see FDA recall press releases.
If you use this product, check the lot, NDC or model number against the official notice and follow its instructions. Do not stop a prescribed medicine or remove a medical device without talking to your doctor or pharmacist. Report problems to FDA MedWatch. FoodRecallNow summarizes public FDA data and is not medical advice.