Buy me a coffee
Home › FDA recalls › IBA Proton Therapy System
CLASS II · ModerateFDA

Ion Beam Applications S.A. recalls IBA Proton Therapy System

📅 Feb 3, 2026 · 📍 Nationwide
Recalling company
Ion Beam Applications S.A.
Product
IBA Proton Therapy System - PROTEUS 235
Reason
It was identified that eh Universal Beam Triggering Interface (UBTI) is disabled in the Therapy Safety System (TSS) configuration. In this condition, the TSS does not interrupt the beam delivery if the Healthy Signal from an external gating device is lost. This may lead to a risk of mistreatment when external gating devices are used.
Classification
Class II – may cause temporary or medically reversible health consequences
Recall date
Feb 3, 2026
Reported
Mar 18, 2026
Distribution
Worldwide - US Nationwide distribution in the state of FL and the countries of Italy and Japan.
Status
Ongoing
Quantity
4 units (1 US, 3 OUS)
Lot / codes
PROTEUS 235; UDI-DI: 05404013801138; Serial Numbers: SBF103 (JP), SBF124 (EU), SBF133 (US), SBF140 (EU)
Company location
Ottignies-Louvain-La-Neuve, N/A
Recall type
Voluntary: Firm initiated
Recall number
Z-1497-2026
Agency
U.S. Food and Drug Administration (FDA)

View official FDA notice ↗ Search recall number Z-1497-2026 in the FDA Enforcement Report. Also see FDA recall press releases.

If you use this product, check the lot, NDC or model number against the official notice and follow its instructions. Do not stop a prescribed medicine or remove a medical device without talking to your doctor or pharmacist. Report problems to FDA MedWatch. FoodRecallNow summarizes public FDA data and is not medical advice.