Staar Surgical AG recalls EVO+ Visian Toric Implantable Collamer Lens
- Recalling company
- Staar Surgical AG
- Product
- EVO+ Visian Toric Implantable Collamer Lens (TICL), Model: VTICM5
- Reason
- Toric Implantable Collamer for correction/reduction of myopia in phakic eyes with astigmatism were labeled as 12.1 mm length but are 12.6 mm length lenses, so there is the potential for excessive vault, which may cause: Significant reduction of irido-corneal angles, Significant shallowing of anterior chamber, Patient discomfort, Elevated IOP/Angle closure, Unreactive pupil, Lens exchange/removal.
- Classification
- Class II – may cause temporary or medically reversible health consequences
- Recall date
- Jan 21, 2026
- Reported
- Mar 11, 2026
- Distribution
- International distribution to the countries of India, Iran, Korea, Saudi Arabia.
- Status
- Ongoing
- Quantity
- 7
- Lot / codes
- UDI-DI: 00840311304363. Expiration Date: 2027-11-30 Serial Numbers: T2286964, T2286971, T2286973, T2286979, T2286980, T2286981, T2286988
- Company location
- Nidau, N/A
- Recall type
- Voluntary: Firm initiated
- Recall number
- Z-1480-2026
- Agency
- U.S. Food and Drug Administration (FDA)
View official FDA notice ↗ Search recall number Z-1480-2026 in the FDA Enforcement Report. Also see FDA recall press releases.
If you use this product, check the lot, NDC or model number against the official notice and follow its instructions. Do not stop a prescribed medicine or remove a medical device without talking to your doctor or pharmacist. Report problems to FDA MedWatch. FoodRecallNow summarizes public FDA data and is not medical advice.