Olympus Corporation of the Americas recalls Brand Name: ShockPulse-SE Lithotripsy System Product Name: ShockPulse-SE Lithotripsy Syst
- Recalling company
- Olympus Corporation of the Americas
- Product
- Brand Name: ShockPulse-SE Lithotripsy System Product Name: ShockPulse-SE Lithotripsy System - Generator Model/Catalog Number: SPL-G Product Description: The ShockPulse Lithotripsy System is an electromechanical device capable of fragmenting calculi and aspirating stone debris. The ShockPulse transducer is a reusable device that requires cleaning and steam sterilization prior to use in order to generate both low-frequency mechanical impacts and high-frequency ultrasonic energy that quickly disintegrates stones.
- Reason
- Ongoing investigations identified additional instances of the device Generator remaining in a blinking phase waiting to recognize the transducer. Damage to the transducer plug and/or generator receptacle may cause these issues.
- Classification
- Class II – may cause temporary or medically reversible health consequences
- Recall date
- Nov 14, 2025
- Reported
- Mar 11, 2026
- Distribution
- Worldwide distribution - US Nationwide and the countries of Canada, Mexico, Bolivia, Chile, India, China, EU, Singapore, Australia, Korea, and Japan.
- Status
- Ongoing
- Quantity
- 1684 units
- Lot / codes
- Model/Catalog Number: SPL-G; UDI: 00821925044203; Serial Numbers: All;
- Company location
- Center Valley, PA
- Recall type
- Voluntary: Firm initiated
- Recall number
- Z-1479-2026
- Agency
- U.S. Food and Drug Administration (FDA)
View official FDA notice ↗ Search recall number Z-1479-2026 in the FDA Enforcement Report. Also see FDA recall press releases.
If you use this product, check the lot, NDC or model number against the official notice and follow its instructions. Do not stop a prescribed medicine or remove a medical device without talking to your doctor or pharmacist. Report problems to FDA MedWatch. FoodRecallNow summarizes public FDA data and is not medical advice.