Buy me a coffee
Home › FDA recalls › Medline Kits containing Tego Connectors Medline DIALYSIS ON
CLASS II · ModerateFDA

Medline Industries, LP recalls Medline Kits containing Tego Connectors Medline DIALYSIS ON/OFF KIT SKU EBSI1746

📅 Jan 8, 2026 · 📍 Nationwide
Recalling company
Medline Industries, LP
Product
Medline Kits containing Tego Connectors Medline DIALYSIS ON/OFF KIT SKU EBSI1746
Reason
Affected lots may exhibit issues with the silicone seal on Tego Connectors included in kits. Issues include silicone seal doming and silicone seal tearing. These issues may result in an occluded fluid path identified by the inability of the user to inject or withdraw blood product with a syringe. Damaged seals may also result in delay in therapy or fluid leakage, interruption in therapy, exposure to biological contaminants or air infused into the body.
Classification
Class II – may cause temporary or medically reversible health consequences
Recall date
Jan 8, 2026
Reported
Mar 4, 2026
Distribution
US Nationwide distribution.
Status
Ongoing
Quantity
2020
Lot / codes
UDI-DI (ea) 10653160373582 UDI-DI (case) 00653160373585 Kit lots 2024051380 2024052080 2024052180 2024052280 2024071580 2025061790 2025072990
Company location
Northfield, IL
Recall type
Voluntary: Firm initiated
Recall number
Z-1464-2026
Agency
U.S. Food and Drug Administration (FDA)

View official FDA notice ↗ Search recall number Z-1464-2026 in the FDA Enforcement Report. Also see FDA recall press releases.

If you use this product, check the lot, NDC or model number against the official notice and follow its instructions. Do not stop a prescribed medicine or remove a medical device without talking to your doctor or pharmacist. Report problems to FDA MedWatch. FoodRecallNow summarizes public FDA data and is not medical advice.