Medline Industries, LP recalls Medline Convenience Kits: 1) SPINAL CDS
- Recalling company
- Medline Industries, LP
- Product
- Medline Convenience Kits: 1) SPINAL CDS, Model Number: CDS940087AG; 2) SPINAL CDS, Model Number: CDS940087AI; 3) TOTAL HIP CDS-LF, Model Number: CDS980832Q; 4) TOTAL HIP CDS-LF, Model Number: CDS980832R; 5) TOTAL HIP CDS-LF, Model Number: CDS980832S; 6) TOTAL KNEE CDS-LF, Model Number: CDS980833Q; 7) TOTAL KNEE CDS-LF, Model Number: CDS980833R; 8) TOTAL HIP PROCEDURE, Model Number: CDS980865V; 9) TOTAL HIP PROCEDURE, Model Number: CDS980865W; 10) TOTAL HIP PROCEDURE, Model Number: CDS980865X; 11) TOTAL HIP PROCEDURE, Model Number: CDS980865Y; 12) TOTAL KNEE CDS, Model Number: CDS981454K; 13) TOTAL KNEE CDS, Model Number: CDS981454L; 14) KNEE ARTHROSCOPY CDS, Model Number: CDS981502K; 15) TOTAL KNEE, Model Number: CDS981505O; 16) TOTAL KNEE, Model Number: CDS981505P; 17) TOTAL KNEE, Model Number: CDS981505Q; 18) TOTAL KNEE, Model Number: CDS981505R; 19) OR LAMINECTOMY CDS, Model Number: CDS981858F; 20) LAMI FUSION, Model Number: CDS982035W; 21) ARMC TOTAL KNEE CDS, Model Number: CDS982281O; 22) ARMC TOTAL KNEE CDS, Model Number: CDS982281P; 23) ARMC TOTAL KNEE CDS, Model Number: CDS982281S; 24) TOTAL HIP CDS, Model Number: CDS983036K; 25) TOTAL KNEE CDS, Model Number: CDS983042K; 26) LOWER EXTREMITY CDS, Model Number: CDS983049K; 27) ZALE SPINE CDS, Model Number: CDS983182G; 28) TOTAL KNEE CDS, Model Number: CDS983297M; 29) SITTING CERVICAL CDS, Model Number: CDS983480G; 30) SHOULDER CDS, Model Number: CDS983581K; 31) SHOULDER ARTHROSCOPY CDS, Model Number: CDS983838G; 32) TOTAL HIP CDS, Model Number: CDS983840I; 33) TOTAL JOINT CDS-LF, Model Number: CDS983863D; 34) SHOULDER ARTHROSCOPY CDS, Model Number: CDS983879C; 35) TOTAL JOINT, Model Number: CDS983903D; 36) GB TOTAL KNEE, Model Number: CDS984147K; 37) GB LOWER EXTREMITY, Model Number: CDS984156L; 38) LOWER EXTREMTIY, Model Number: CDS984254K; 39) LOWER EXTREMTIY, Model Number: CDS984254L; 40) LOWER EXTREMTIY, Model Number: CDS984254M; 41) LAMI MICRODISC-LF, Model Number: CDS984527I; 42) EXTREMITY CDS-LF, Model Number: CDS984607G; 43) EXTREMITY CDS-LF, Model Number: CDS984607I; 44) HANA/FRACTURE TABLE CDS-LF, Model Number: CDS984610C; 45) TOTAL JOINT CDS, Model Number: CDS985202B; 46) TOTAL JOINT CDS, Model Number: CDS985202C; 47) TRAUMA-LF, Model Number: CDS985279K; 48) ARTHROSCOPY CDS, Model Number: CDS985316A; 49) EXTREMITY CDS, Model Number: CDS985317C; 50) EXTREMITY CDS, Model Number: CDS985317D; 51) EXTREMITY CDS, Model Number: CDS985317F; 52) EXTREMITY CDS, Model Number: CDS985317G; 53) KNEE ARTHROSCOPY CDS, Model Number: CDS985425J; 54) TOTAL HIP PACK (CDSOPM834)-LF, Model Number: CDSOPM834W; 55) JOINT ASPIRATION KIT, Model Number: DYKM1161; 56) TOTAL KNEE SUPPLEMENT PACK NS, Model Number: DYKM2106; 57) KIT ORTHRO TOTAL HIP DR. NANDI, Model Number: DYKMBNDL137; 58) KIT NEUR FUS LUM POST CRANDALL, Model Number: DYKMBNDL138A; 59) KIT NEURO CRANIOTOMYANEURYSM R, Model Number: DYKMBNDL141; 60) KIT NEURO CRANIOTOMYANEURYSM R, Model Number: DYKMBNDL141A; 61) KIT NEURO CRANIOTOMYANEURYSM R, Model Number: DYKMBNDL141B; 62) KIT NEURO FUS LUM POST SANSUR, Model Number: DYKMBNDL142C; 63) KIT NEURO CRANIOTOMY ACOUSTIC, Model Number: DYKMBNDL143; 64) KIT NEURO CRANIOTOMY FOR TUMOR, Model Number: DYKMBNDL144; 65
- Reason
- Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.
- Classification
- Class II – may cause temporary or medically reversible health consequences
- Recall date
- Jan 7, 2026
- Reported
- Feb 25, 2026
- Distribution
- Worldwide distribution - US Nationwide.
- Status
- Ongoing
- Quantity
- 117907 units
- Lot / codes
- 1) CDS940087AG, UDI-DI: 10193489847260(each), 40193489847261(case), Lot Number: 21EBC900; 2) CDS940087AI, UDI-DI: 10193489933758(each), 40193489933759(case), Lot Number: 22GBE442; 3) CDS940087AI, UDI-DI: 10193489933758(each), 40193489933759(case), Lot Number: 22EBG548; 4) CDS940087AI, UDI-DI: 10193489933758(each), 40193489933759(case), Lot Number: 22EBF695; 5) CDS940087AI, UDI-DI: 10193489933758(each), 40193489933759(case), Lot Number: 22EBA274; 6) CDS940087AI, UDI-DI: 10193489933758(each), 40193489933759(case), Lot Number: 22BBY060; 7) CDS940087AI, UDI-DI: 10193489933758(each), 40193489933759(case), Lot Number: 21KBH163; 8) CDS940087AI, UDI-DI: 10193489933758(each), 40193489933759(case), Lot Number: 21JBF936; 9) CDS940087AI, UDI-DI: 10193489933758(each), 40193489933759(case), Lot Number: 21IBJ260; 10) CDS940087AI, UDI-DI: 10193489933758(each), 40193489933759(case), Lot Number: 21GBI183; 11) CDS980832Q, UDI-DI: 10193489788655(each), 40193489788656(case), Lot Number: 21FBS254; 12) CDS980832Q, UDI-DI: 10193489788655(each), 40193489788656(case), Lot Number: 21BBL223; 13) CDS980832R, UDI-DI: 10193489900972(each), 40193489900973(case), Lot Number: 21LBO219; 14) CDS980832R, UDI-DI: 10193489900972(each), 40193489900973(case), Lot Number: 21KBQ650; 15) CDS980832R, UDI-DI: 10193489900972(each), 40193489900973(case), Lot Number: 21KBB625; 16) CDS980832R, UDI-DI: 10193489900972(each), 40193489900973(case), Lot Number: 21IBV695; 17) CDS980832R, UDI-DI: 10193489900972(each), 4019348990097
- Company location
- Northfield, IL
- Recall type
- Voluntary: Firm initiated
- Recall number
- Z-1421-2026
- Agency
- U.S. Food and Drug Administration (FDA)
View official FDA notice ↗ Search recall number Z-1421-2026 in the FDA Enforcement Report. Also see FDA recall press releases.
If you use this product, check the lot, NDC or model number against the official notice and follow its instructions. Do not stop a prescribed medicine or remove a medical device without talking to your doctor or pharmacist. Report problems to FDA MedWatch. FoodRecallNow summarizes public FDA data and is not medical advice.