Buy me a coffee
Home › FDA recalls › Medline Convenience Kits: 1) BAP NORTH ROBOTIC SURGERY CDS
CLASS II · ModerateFDA

Medline Industries, LP recalls Medline Convenience Kits: 1) BAP NORTH ROBOTIC SURGERY CDS

📅 Jan 7, 2026 · 📍 Nationwide
Recalling company
Medline Industries, LP
Product
Medline Convenience Kits: 1) BAP NORTH ROBOTIC SURGERY CDS, Model Number: CDS982407N; 2) TOT/ GYN LITHOTOMY CDS, Model Number: CDS985431K; 3) TOT/ GYN LITHOTOMY CDS, Model Number: CDS985431O; 4) KIT ROBOTICS GYN, Model Number: DYKMBNDL109A; 5) KIT ROBOTICS UROLOGY PROSTATE, Model Number: DYKMBNDL116A; 6) KIT ROBOTICS UROLOGY PROSTATE, Model Number: DYKMBNDL116G; 7) ROBOTIC PACK, Model Number: DYNJ04135B; 8) MINOR LITHOTOMY PACK, Model Number: DYNJ04382J; 9) NBMC D & C PACK-LF, Model Number: DYNJ0665843C; 10) D&C PACK-LF, Model Number: DYNJ22567G; 11) D&C PACK-LF, Model Number: DYNJ22567J; 12) D AND C, Model Number: DYNJ41272B; 13) D AND C, Model Number: DYNJ41272C; 14) OB PACK, Model Number: DYNJ42884; 15) D&C/GYN PACK, Model Number: DYNJ47713A; 16) PERI GYN PACK, Model Number: DYNJ55377D; 17) ANTERIOR CERVICAL-LF, Model Number: DYNJ58344; 18) D&C PACK, Model Number: DYNJ67214D; 19) D&C PACK, Model Number: DYNJ67214F; 20) D&C PACK, Model Number: DYNJ67214G; 21) D&C PACK, Model Number: DYNJ67214I; 22) LITHOTOMY PACK, Model Number: DYNJ80765C; 23) GYN PACK, Model Number: DYNJ83475; 24) D AND C PACK, Model Number: DYNJ87468; 25) D&C-ASC, Model Number: DYNJ900465C; 26) D&C-ASC, Model Number: DYNJ900465D; 27) D&C-ASC, Model Number: DYNJ900465F; 28) GYN LITHOTOMY-LF, Model Number: DYNJ902711L; 29) GYN LITHOTOMY-LF, Model Number: DYNJ902711N; 30) D&C, Model Number: DYNJ906947C; 31) PACK PERI GYN DSMC, Model Number: DYNJT6424
Reason
Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.
Classification
Class II – may cause temporary or medically reversible health consequences
Recall date
Jan 7, 2026
Reported
Feb 25, 2026
Distribution
Worldwide distribution - US Nationwide.
Status
Ongoing
Quantity
7725 units
Lot / codes
1) CDS982407N, UDI-DI: 10193489683004(each), 40193489683005(case), Lot Number: 21DBO201; 2) CDS982407N, UDI-DI: 10193489683004(each), 40193489683005(case), Lot Number: 21CBD150; 3) CDS985431K, UDI-DI: 10195327451295(each), 40195327451296(case), Lot Number: 23GBM332; 4) CDS985431K, UDI-DI: 10195327451295(each), 40195327451296(case), Lot Number: 23GBL932; 5) CDS985431O, UDI-DI: 10198459195006(each), 40198459195007(case), Lot Number: 25GBB847; 6) DYKMBNDL109A, UDI-DI: 10193489848779(each), 40193489848770(case), Lot Number: 21IBS290; 7) DYKMBNDL109A, UDI-DI: 10193489848779(each), 40193489848770(case), Lot Number: 21GBE967; 8) DYKMBNDL109A, UDI-DI: 10193489848779(each), 40193489848770(case), Lot Number: 21FBB105; 9) DYKMBNDL109A, UDI-DI: 10193489848779(each), 40193489848770(case), Lot Number: 21DBC065; 10) DYKMBNDL109A, UDI-DI: 10193489848779(each), 40193489848770(case), Lot Number: 21CBQ508; 11) DYKMBNDL116A, UDI-DI: 10193489880717(each), 40193489880718(case), Lot Number: 22HBE127; 12) DYKMBNDL116A, UDI-DI: 10193489880717(each), 40193489880718(case), Lot Number: 22GBL821; 13) DYKMBNDL116A, UDI-DI: 10193489880717(each), 40193489880718(case), Lot Number: 22FBM235; 14) DYKMBNDL116A, UDI-DI: 10193489880717(each), 40193489880718(case), Lot Number: 22FBC878; 15) DYKMBNDL116A, UDI-DI: 10193489880717(each), 40193489880718(case), Lot Number: 22EBP063; 16) DYKMBNDL116A, UDI-DI: 10193489880717(each), 40193489880718(case), Lot Number: 22EBP060; 17) DYKMBNDL116A, UDI-DI: 10193489880717(each),
Company location
Northfield, IL
Recall type
Voluntary: Firm initiated
Recall number
Z-1419-2026
Agency
U.S. Food and Drug Administration (FDA)

View official FDA notice ↗ Search recall number Z-1419-2026 in the FDA Enforcement Report. Also see FDA recall press releases.

If you use this product, check the lot, NDC or model number against the official notice and follow its instructions. Do not stop a prescribed medicine or remove a medical device without talking to your doctor or pharmacist. Report problems to FDA MedWatch. FoodRecallNow summarizes public FDA data and is not medical advice.