Olympus Corporation of the Americas recalls Olympus Thunderbeat
- Recalling company
- Olympus Corporation of the Americas
- Product
- Olympus Thunderbeat, 5 mm, 45 cm, Pistol Grip
- Reason
- Firm is initiating a removal due to continued reports of adverse events.
- Classification
- Class II – may cause temporary or medically reversible health consequences
- Recall date
- Oct 22, 2025
- Reported
- Feb 25, 2026
- Distribution
- Worldwide distribution - US Nationwide and the countries of Brazil, Canada, Germany, Mexico, and Japan.
- Status
- Ongoing
- Quantity
- 30 units
- Lot / codes
- Model No. TB-0545PC; UDI: 04953170308635; All Lots.
- Company location
- Center Valley, PA
- Recall type
- Voluntary: Firm initiated
- Recall number
- Z-1377-2026
- Agency
- U.S. Food and Drug Administration (FDA)
View official FDA notice ↗ Search recall number Z-1377-2026 in the FDA Enforcement Report. Also see FDA recall press releases.
If you use this product, check the lot, NDC or model number against the official notice and follow its instructions. Do not stop a prescribed medicine or remove a medical device without talking to your doctor or pharmacist. Report problems to FDA MedWatch. FoodRecallNow summarizes public FDA data and is not medical advice.