Buy me a coffee
Home › FDA recalls › Laerdal Compact Suction Unit 4 RTCA version
CLASS II · ModerateFDA

Laerdal Medical Corporation recalls Laerdal Compact Suction Unit 4 RTCA version

📅 Jan 13, 2026 · 📍 Nationwide
Recalling company
Laerdal Medical Corporation
Product
Laerdal Compact Suction Unit 4 RTCA version (LCSU 4 RTCA) 1) LCSU 4, 800 ml, RTCA; Catalog Number: 880052; 2) LCSU 4, 300 ml, RTCA; Catalog Number: 880062; 3) LCSU 4, RTCA; Catalog Number: 881152;
Reason
units manufactured in a limited time period may emit electromagnetic noise beyond the acceptable limits for RTCA application specified in RTCA DO-160G Chapter 21, Equipment Category M.
Classification
Class II – may cause temporary or medically reversible health consequences
Recall date
Jan 13, 2026
Reported
Feb 18, 2026
Distribution
Worldwide distribution - US Nationwide and the countries of Denmark, Sweden, Norway, Canada, Finland, Australia, Spain, France, United Kingdom, Italy, Korea, Germany, Netherlands, Switzerland, New Zealand, Singapore, Malaysia, Austria.
Status
Ongoing
Quantity
1202 units (537 US, 665 OUS) (2/23/2026: 574 US, 1239 units)
Lot / codes
1) LCSU 4, 800 ml, RTCA; Catalog Number: 880052; UDI-DI: 07045432088519; 2) LCSU 4, 300 ml, RTCA; Catalog Number: 880062; UDI-DI: 07045432088533; 3) LCSU 4, RTCA; Catalog Number: 881152; UDI-DI: 07045432067132;
Company location
Wappingers Falls, NY
Recall type
Voluntary: Firm initiated
Recall number
Z-1341-2026
Agency
U.S. Food and Drug Administration (FDA)

View official FDA notice ↗ Search recall number Z-1341-2026 in the FDA Enforcement Report. Also see FDA recall press releases.

If you use this product, check the lot, NDC or model number against the official notice and follow its instructions. Do not stop a prescribed medicine or remove a medical device without talking to your doctor or pharmacist. Report problems to FDA MedWatch. FoodRecallNow summarizes public FDA data and is not medical advice.