Laerdal Medical Corporation recalls Laerdal Compact Suction Unit 4 RTCA version
- Recalling company
- Laerdal Medical Corporation
- Product
- Laerdal Compact Suction Unit 4 RTCA version (LCSU 4 RTCA) 1) LCSU 4, 800 ml, RTCA; Catalog Number: 880052; 2) LCSU 4, 300 ml, RTCA; Catalog Number: 880062; 3) LCSU 4, RTCA; Catalog Number: 881152;
- Reason
- units manufactured in a limited time period may emit electromagnetic noise beyond the acceptable limits for RTCA application specified in RTCA DO-160G Chapter 21, Equipment Category M.
- Classification
- Class II – may cause temporary or medically reversible health consequences
- Recall date
- Jan 13, 2026
- Reported
- Feb 18, 2026
- Distribution
- Worldwide distribution - US Nationwide and the countries of Denmark, Sweden, Norway, Canada, Finland, Australia, Spain, France, United Kingdom, Italy, Korea, Germany, Netherlands, Switzerland, New Zealand, Singapore, Malaysia, Austria.
- Status
- Ongoing
- Quantity
- 1202 units (537 US, 665 OUS) (2/23/2026: 574 US, 1239 units)
- Lot / codes
- 1) LCSU 4, 800 ml, RTCA; Catalog Number: 880052; UDI-DI: 07045432088519; 2) LCSU 4, 300 ml, RTCA; Catalog Number: 880062; UDI-DI: 07045432088533; 3) LCSU 4, RTCA; Catalog Number: 881152; UDI-DI: 07045432067132;
- Company location
- Wappingers Falls, NY
- Recall type
- Voluntary: Firm initiated
- Recall number
- Z-1341-2026
- Agency
- U.S. Food and Drug Administration (FDA)
View official FDA notice ↗ Search recall number Z-1341-2026 in the FDA Enforcement Report. Also see FDA recall press releases.
If you use this product, check the lot, NDC or model number against the official notice and follow its instructions. Do not stop a prescribed medicine or remove a medical device without talking to your doctor or pharmacist. Report problems to FDA MedWatch. FoodRecallNow summarizes public FDA data and is not medical advice.