Buy me a coffee
Home › FDA recalls › Brilliance iCT
CLASS II · ModerateFDA

Philips North America Llc recalls Brilliance iCT

📅 Jan 7, 2026 · 📍 Nationwide
Recalling company
Philips North America Llc
Product
Brilliance iCT; Product Code (REF): 728306;
Reason
As part of the CT systems sub assembly, thirty-two (32) fasteners are used to attach the rotor to the bearing within the gantry. One or more of these fasteners may not be torqued to specification. If multiple fasteners are not torqued to specification, the rotor or rotor parts may become unsecured or displaced. While there has been no reported or observed event of expelled parts during gantry rotation, unsecured rotor parts may potentially be expelled.
Classification
Class II – may cause temporary or medically reversible health consequences
Recall date
Jan 7, 2026
Reported
Feb 18, 2026
Distribution
Worldwide - US Nationwide distribution in the states of AL, AR, CA, CO, CT, FL, GA, HI, IL, IN, KY, LA, MA, MD, MN, NC, NY, OH, OR, PA, TX, VT, WA, WV and the countries of Australia, Austria, Bahrain, Belgium, Brazil, Bulgaria, Chile, China, Costa Rica, Croatia, Czech Republic, Denmark, Ecuador, Egypt, El Salvador, France, Germany, Hong Kong, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kuwait, Lithuania, Macao, Mexico, Netherlands, New Zealand, Norway, Panama, Philippines, Poland, Portugal, R¿union, Russian Fed., Saudi Arabia, Serbia, Singapore, South Africa, South Korea, Spain, Switzerland, Taiwan, Thailand, Turkmenistan, United Kingdom, Uzbekistan, Vietnam.
Status
Ongoing
Quantity
47 units
Lot / codes
Product Code (REF): 728306; UDI-DI: 00884838059474; Serial Numbers: 87064, 87120, 87121, 87122, 87125, 87126, 87127, 87128, 87129, 87130, 87135, 87143, 87148, 87152, 87153, 87157, 87160, 87162, 87167, 87173, 87174, 87176, 87181, 87197, 87198, 87199, 87200 , 87202, 87203, 87204, 87205, 87206, 87207, 87208, 87209, 87210, 87211, 87212, 87213 , 87214, 87215, 87216, 87217, 87218, 87219, 87220, 87221;
Company location
Cambridge, MA
Recall type
Voluntary: Firm initiated
Recall number
Z-1315-2026
Agency
U.S. Food and Drug Administration (FDA)

View official FDA notice ↗ Search recall number Z-1315-2026 in the FDA Enforcement Report. Also see FDA recall press releases.

If you use this product, check the lot, NDC or model number against the official notice and follow its instructions. Do not stop a prescribed medicine or remove a medical device without talking to your doctor or pharmacist. Report problems to FDA MedWatch. FoodRecallNow summarizes public FDA data and is not medical advice.