Buy me a coffee
Home › FDA recalls › Brand Name: Molift Product Name: Molift Mover 205
CLASS II · ModerateFDA

ETAC A/S recalls Brand Name: Molift Product Name: Molift Mover 205

📅 Nov 17, 2025 · 📍 Nationwide
Recalling company
ETAC A/S
Product
Brand Name: Molift Product Name: Molift Mover 205 Model/Catalog Number: (1) M14000, (2) M1401, (3) M1402, (4) M14050, (5) M14150, (6) M1402-01, (7) M14050-00 (8) M14001;
Reason
A customer experienced that during assembly and load testing of the device, the bolt for mounting the lifting bar broke. Upon further investigation of this customer complaint, it became clear that the defect occurs as a result of deformation of the materials involved between the lifting bar and the arm of the device. Such deformation can lead to limited mobility in the joint, which prevents proper functionality of the equipment.
Classification
Class II – may cause temporary or medically reversible health consequences
Recall date
Nov 17, 2025
Reported
Feb 11, 2026
Distribution
Worldwide - US Nationwide distribution in the state of PA and the countries of Australia, Czech Republic, Finland, France, Germany, Israel, Italy, Japan, New Zealand, Norway, Poland, Qatar Spain, Sweden, Switzerland, UK.
Status
Ongoing
Quantity
95 units
Lot / codes
(1) Model/Catalog Number: M14000; UDI-DI: 05707995032270; Serial Numbers: 616606, 616604, 591003, 591002, 590906, 590901, 590904, 590903, 590902, 591001, 604101, 604005, 591703, 591007, 591006, 591005, 591004, 615105, 615104, 615103, 615102, 615101, 604004, 604003, 604002, 604001, 614202, 614201, 591702, 591701, 616603, 615106, 590905; (2) Model/Catalog Number: M1401; UDI-DI: 05707995032270; Serial Numbers: 593302, 593301, 593303, 612507, 612506, 612508, 612505, 612503, 620702, 612504, 612502, 612501, 593304, 593305, 620703, 620701; (3) Model/Catalog Number: M1402; UDI-DI: 05707995032270; Serial Numbers: 612601; (4) Model/Catalog Number: M14050; UDI-DI: 05707995032270; Serial Numbers: 590801, 596701, 585701, 584501, 591801, 604405, 604406, 604402, 604404, 604403, 604401, 604303, 604302, 604301; (5) Model/Catalog Number: M14150; UDI-DI: 05707995032270; Serial Numbers: 585811, 585810, 585809, 585808, 585806, 585805, 585804, 585803, 585801, 596604, 596603, 596602, 596601, 596611, 596610, 596609, 596608, 596607, 596606, 596605, 585807, 585802; (6) Model/Catalog Number: M1402-01; UDI-DI: 05707995032270; Serial Numbers: 612701, 612703, 612702; (7) Model/Catalog Number: M14050-00; UDI-DI: 05707995032270; Serial Numbers: 584307; (8) Model/Catalog Number: M14001; UDI-DI: 05707995032270; Serial Numbers: 603501, 599801, 599802, 604202, 604201;
Company location
Gedved, N/A
Recall type
Voluntary: Firm initiated
Recall number
Z-1308-2026
Agency
U.S. Food and Drug Administration (FDA)

View official FDA notice ↗ Search recall number Z-1308-2026 in the FDA Enforcement Report. Also see FDA recall press releases.

If you use this product, check the lot, NDC or model number against the official notice and follow its instructions. Do not stop a prescribed medicine or remove a medical device without talking to your doctor or pharmacist. Report problems to FDA MedWatch. FoodRecallNow summarizes public FDA data and is not medical advice.