Buy me a coffee
Home › FDA recalls › Bubble Sensor for 3/8" x 3/32" tubing
CLASS II · ModerateFDA

Maquet Cardiopulmonary Gmbh recalls Bubble Sensor for 3/8" x 3/32" tubing

📅 Jan 9, 2026 · 📍 Nationwide
Recalling company
Maquet Cardiopulmonary Gmbh
Product
Bubble Sensor for 3/8" x 3/32" tubing; Model Number: 701055720;
Reason
Internal investigations have identified an issue with the durability of the connecting cable near the connection to the Venous Bubble Sensor. Excessive bending of the connecting cable can lead to full damage of the Venous Bubble Sensor or loose cable contact, which may trigger the errors Ven. bubble sensor defective or Ven. bubble sensor disconnected on the connected medical device. These errors can occur temporarily when the connecting cable is moved or permanently if the connection is fully compromised.
Classification
Class II – may cause temporary or medically reversible health consequences
Recall date
Jan 9, 2026
Reported
Feb 11, 2026
Distribution
Worldwide Distribution: US (Nationwide) and OUS (International) to countries of: Albania, Algeria, Argentina, Australia, Austria, Azerbaijan, Bahrain, Belgium, Bosnia and Herzegovina, Botswana, Brazil, Bulgaria, Canada, Chile, China, Colombia, Costa Rica, Croatia, Cyprus, Czech Republic, Denmark, El Salvador, Estonia, Finland, France, Germany, Hong Kong, Hungary, India, Indonesia, Iraq, Ireland, Israel, Italy, Japan, Korea, Republic of (South Korea), Latvia, Lithuania, Mexico, Morocco, Namibia, Netherlands, New Zealand, Nicaragua, Norway, Pakistan, Philippines, Poland, Portugal, Qatar, Romania, Russian Federation, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, Turkmenistan, Ukraine, United Arab Emirates, United Kingdom, United States, Uruguay, Uzbekistan, and Vietnam
Status
Ongoing
Quantity
3050 units
Lot / codes
Model Number: 701055720; UDI-DI: 04037691816432; Serial Numbers: All units impacted;
Company location
Rastatt, N/A
Recall type
Voluntary: Firm initiated
Recall number
Z-1300-2026
Agency
U.S. Food and Drug Administration (FDA)

View official FDA notice ↗ Search recall number Z-1300-2026 in the FDA Enforcement Report. Also see FDA recall press releases.

If you use this product, check the lot, NDC or model number against the official notice and follow its instructions. Do not stop a prescribed medicine or remove a medical device without talking to your doctor or pharmacist. Report problems to FDA MedWatch. FoodRecallNow summarizes public FDA data and is not medical advice.