Home › FDA recalls › FDR Visionary Suite
CLASS II · ModerateFDA

FUJIFILM Healthcare Americas Corporation recalls FDR Visionary Suite

📅 Jan 9, 2026 · 📍 Nationwide
Recalling company
FUJIFILM Healthcare Americas Corporation
Product
FDR Visionary Suite; Model Number: CH-200; Version: (1) 566-16130-23, (2) 566-16130-31, (3)
Reason
It was found that the contacts of the electromagnetic contactors used inside the cabinet of the X-ray high-voltage generator were welded, and the current might continue to be supplied to the power supply circuit even if the power of the device is cut off. As a result, it cannot be denied that the circuit in question overheats and damages the surrounding parts. It may also cause smoke.
Classification
Class II – may cause temporary or medically reversible health consequences
Recall date
Jan 9, 2026
Reported
Feb 11, 2026
Distribution
US Nationwide distribution in the states of AK, CA, CT, FL, IL, NH, OH, WI.
Status
Ongoing
Quantity
16 units
Lot / codes
Model Number: CH-200; (1) Version: 566-16130-23; UDI-DI: 04540217052226; Serial numbers: MP95A9F6A001, MP95A9F6B001, MP95A9F74001, MP95A9F74002, MP95A9F76001, MP95A9F78001; (2) Version: 566-16130-31; UDI-DI: 04540217057436; Serial numbers: MP95AA783001; (3) Version: 566-16130-33; UDI-DI: 04540217057450; Serial numbers: MP95AA983001, MP95AA983002, MP95AA983003, MP95AA985001, MP95AA985003, MP95AA985002, MP95AA988001, MP95AA988002;
Company location
Lexington, MA
Recall type
Voluntary: Firm initiated
Recall number
Z-1299-2026
Agency
U.S. Food and Drug Administration (FDA)

View official FDA notice ↗ Search recall number Z-1299-2026 in the FDA Enforcement Report. Also see FDA recall press releases.

If you use this product, check the lot, NDC or model number against the official notice and follow its instructions. Do not stop a prescribed medicine or remove a medical device without talking to your doctor or pharmacist. Report problems to FDA MedWatch. FoodRecallNow summarizes public FDA data and is not medical advice.