NIPRO Technical Services, Inc. recalls Conductivity Standard Solution

📅 Dec 12, 2025 · 📍 Nationwide
Recalling company
NIPRO Technical Services, Inc.
Product
Conductivity Standard Solution Catalog/Model Numbers: 10001, 10011, 10021, 10032, 10040, 10040c, and 10041 Conductivity Standard Solutions are a secondary standard solution used for the calibration of conductivity cells together with conductivity meters
Reason
due to a defect associated with the conductivity values of the solution used to calibrate meters and result in the reporting of incorrect conductivity values for dialysate.
Classification
Class II – may cause temporary or medically reversible health consequences
Recall date
Dec 12, 2025
Reported
Feb 4, 2026
Distribution
U.S. Nationwide distribution in the states of AL, AR, AZ, CA, DE, FL, GA. IL, IN, KS, MI, MO, NC, ND, NJ, NV, NY, OH, OK, OR, PA, SC, TN TX, UT, VA, and WA.
Status
Ongoing
Quantity
499 units
Lot / codes
Catalog/Model Numbers/ Lot Numbers / UDI codes: 10001 14250616-1019106 (01) 0 0850016 97600 0 (17) 251030 (10) SHOWCASE 10011 14250616-1019107 (01) 0 0850016 97609 3 (17) 251030 (10) SHOWCASE 10040c 50250624-1019120 (01) 0 0850016 97636 9 (17) 251030 (10) SHOWCASE 10040 50250624-1019119 (01) 0 0850016 97635 2 (17) 251030 (10) SHOWCASE 10032 14250616-1019108 (01) 0 0850016 97631 1 (17) 251030 (10) SHOWCASE 10041 50250522-1018643 (01) 0 0850016 97637 6 (17) 251030 (10) SHOWCASE 10041 50250624-1019121 (01) 0 0850016 97637 6 (17) 251030 (10) SHOWCASE 10021 15F250603-1018834 (01) 0 0850016 97619 2 (17) 251030 (10) SHOWCASE
Company location
Mesa, AZ
Recall type
Voluntary: Firm initiated
Recall number
Z-1185-2026
Agency
U.S. Food and Drug Administration (FDA)

View official FDA notice ↗ Search recall number Z-1185-2026 in the FDA Enforcement Report. Also see FDA recall press releases.

If you use this product, check the lot, NDC or model number against the official notice and follow its instructions. Do not stop a prescribed medicine or remove a medical device without talking to your doctor or pharmacist. Report problems to FDA MedWatch. FoodRecallNow summarizes public FDA data and is not medical advice.