Home › FDA recalls › Alphenix INFX-8000H
CLASS II · ModerateFDA

Canon Medical System, USA, INC. recalls Alphenix INFX-8000H

📅 Dec 19, 2025 · 📍 Nationwide
Recalling company
Canon Medical System, USA, INC.
Product
Alphenix INFX-8000H, interventional fluoroscopic x-ray system
Reason
It has been discovered that there is a possibility that the fixing screws of the ceiling movement gear in the ceiling suspended arm that are part of the X-ray interventional systems may become loose. As a result, ceiling lateral movement may no longer be possible, abnormal noise may be generated from the suspension system and a sensor error may be displayed
Classification
Class II – may cause temporary or medically reversible health consequences
Recall date
Dec 19, 2025
Reported
Feb 4, 2026
Distribution
Worldwide distribution - US Nationwide and the country of Dominican Republic.
Status
Ongoing
Quantity
3 units
Lot / codes
serial numbers: D2A1672385, A2A20Z2165, B2A1692426.
Company location
Tustin, CA
Recall type
Voluntary: Firm initiated
Recall number
Z-1178-2026
Agency
U.S. Food and Drug Administration (FDA)

View official FDA notice ↗ Search recall number Z-1178-2026 in the FDA Enforcement Report. Also see FDA recall press releases.

If you use this product, check the lot, NDC or model number against the official notice and follow its instructions. Do not stop a prescribed medicine or remove a medical device without talking to your doctor or pharmacist. Report problems to FDA MedWatch. FoodRecallNow summarizes public FDA data and is not medical advice.