Buy me a coffee
Home › FDA recalls › Interventional Fluoroscopic X-Ray System: LUMINOS Q.namix R
CLASS II · ModerateFDA

Siemens Medical Solutions USA, Inc recalls Interventional Fluoroscopic X-Ray System: LUMINOS Q.namix R

📅 Dec 29, 2025 · 📍 Nationwide
Recalling company
Siemens Medical Solutions USA, Inc
Product
Interventional Fluoroscopic X-Ray System: LUMINOS Q.namix R; Siemens Material Number (SMN): 11574002; With Software Version VB10D-SP02
Reason
Ortho images acquired with preset automatic horizontal flip are not flipped and when acquiring single images, there is the possibility that an image of a previous patient may be processed.
Classification
Class II – may cause temporary or medically reversible health consequences
Recall date
Dec 29, 2025
Reported
Feb 4, 2026
Distribution
Worldwide - US Nationwide distribution in the states of AK, CA, IL, MN, NY, OR, VA, WI and the countries of Austria, Belgium, Denmark, France, Germany, Greece, Ireland, Italy, Netherlands, Norway, South Africa, Spain, Sweden, Switzerland.
Status
Ongoing
Quantity
41 units
Lot / codes
Siemens Material Number (SMN): 11574002; UDI-DI: 04056869993973; Serial Numbers: 150050, 150035, 150033, 150034, 150048, 150021, 150020, 150053, 150023, 150044, 150049, 150037, 150054, 150011, 150059, 150024, 150038, 150051, 150029, 150045, 150032, 150056, 150043, 150055, 150041, 150026, 150025, 150022, 150031, 150039, 150046, 150042, 150028, 150047, 150030, 150052, 150027, 150057, 150040, 150058, 150063;
Company location
Malvern, PA
Recall type
Voluntary: Firm initiated
Recall number
Z-1174-2026
Agency
U.S. Food and Drug Administration (FDA)

View official FDA notice ↗ Search recall number Z-1174-2026 in the FDA Enforcement Report. Also see FDA recall press releases.

If you use this product, check the lot, NDC or model number against the official notice and follow its instructions. Do not stop a prescribed medicine or remove a medical device without talking to your doctor or pharmacist. Report problems to FDA MedWatch. FoodRecallNow summarizes public FDA data and is not medical advice.