Boston Scientific Corporation recalls AXIOS Stent and Electrocautery-Enhanced Delivery System
- Recalling company
- Boston Scientific Corporation
- Product
- AXIOS Stent and Electrocautery-Enhanced Delivery System - 8mm x 8mm; Material Number (UPN): M00553690;
- Reason
- Increased reports of stent deployment and expansion issues. The most common outcome reasonably foreseeable resulting from difficulty deploying the stent is prolongation of the procedure to exchange the device for a new one. The most serious reasonably foreseeable outcome in cases where the first flange is unable to deploy or expand would be a cascade of events resulting in additional intervention, either endoscopic or surgical, to remove the stent and close the puncture site.
- Classification
- Class I – a reasonable probability that use will cause serious adverse health consequences or death
- Recall date
- Dec 19, 2025
- Reported
- Feb 11, 2026
- Distribution
- Worldwide distribution. US Nationwide including Puerto Rico; Algeria, Argentina, Austria, Belgium, Brazil, Canada, CHILE, Colombia, Costa Rica, Croatia, Czech Republic, Denmark, Finland, France, Germany, Greece, Hungary, India, Indonesia, Iran, Ireland, Israel, Italy, JAPAN, Jordan, Kuwait, Lebanon, Luxembourg, Malaysia, Mexico, Morocco, Netherlands, Norway, Oman, Pakistan, Palestine, Panama, Paraguay, PERU, Poland, Portugal, Romania, Saudi Arabia, Serbia, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, UAE, and United Kingdom.
- Status
- Ongoing
- Quantity
- 59 units
- Lot / codes
- Material Number (UPN): M00553690; UDI-DI: 191506008093; Lot Numbers: 35960775, 35960777, 35962525, 36100561, 36101748, 36111316, 36154532, 36154533, 36241501, 36251362, 36253016, 36293839, 36301625, 36303192, 36420219, 36425888, 36449561, 36458830, 36461449, 36478580, 36492661, 36495673, 36573597, 36573598, 36573599, 36574152, 36576474, 36584299, 36604665, 36731807, 36731808, 36737997, 36743059, 36743200, 36904231, 36904232, 36904233, 36906587, 36907000, 36907001, 36907032, 37217612, 37217613, 37217614, 37440330, 37495327, 37511242, 37602825, 37602826, 37689946, 37694053, 37694054, 37764851, 37764852, 37995048, 37995049, 37995050, 38025042, 38025043,
- Company location
- Marlborough, MA
- Recall type
- Voluntary: Firm initiated
- Recall number
- Z-1160-2026
- Agency
- U.S. Food and Drug Administration (FDA)
View official FDA notice ↗ Search recall number Z-1160-2026 in the FDA Enforcement Report. Also see FDA recall press releases.
If you use this product, check the lot, NDC or model number against the official notice and follow its instructions. Do not stop a prescribed medicine or remove a medical device without talking to your doctor or pharmacist. Report problems to FDA MedWatch. FoodRecallNow summarizes public FDA data and is not medical advice.